12.2 RS232 troubleshooting

Problem
No connection via RS232
(wrong cable)
No connection via RS232
(cable defect)
No connection via RS232
(wrong COM port)
No connection via RS232
(wrong settings)
No connection via RS232
(blocked COM port)
Acceleration via external device too slow.
Any other problem

12.3 Error messages

Following error messages may be displayed on the UserTerminal:
Error code
Acoustic code
(x = short beep,
o = long beep)
E01
ooooo xoooo
E02
ooooo xxooo
E20
xxooo ooooo
E21
xxooo xoooo
E30
xxxoo ooooo
E31
xxxoo xoooo
E32
xxxoo xxooo
E41
xxxxo xoooo
E50
xxxxx ooooo
E51
xxxxx xoooo
E52
xxxxx xxooo
Solution
For connection to PC with h/p/cosmos software and most external devices
use the included RS 232 interface connection cable [cos00097010034].
Check cable and plugs for defects.
Replace defective cable.
Do not use COM 3 on the device
(see "connection of external devices").
Choose correct interface protocol on device ("User Options" OP 20 or 21).
Choose correct interface protocol on periphery.
Check installation of peripheral software.
Restart peripheral software.
Restart peripheral device.
Check max. acceleration and deceleration level via RS232
("User Options" OP 29).
unplug device, exclude reconnection, mark clearly and inform h/p/cosmos
service personnel via telephone and writing.
Error message
Oil Help
Service Help
Elev Help
Incr Help
Setup Help
Setup Help
Setup Help
Setup Help
FU Help
FU Help
FU Help
Action
See "lubrication"
See "safety inspection"
Unplug device,
exclude reconnection and
contact
service@h-p-cosmos.com