If the heart rate signal totally fails, an acoustic warning signal occurs and the heart rate display shows no value any more.
Furthermore, the device reduces speed and elevation to 0 within one minute.
For possibilities to interfere with an automatic program, see "interfere with an automatic program".
The treadmill adjusts speed and elevation according to following matrices.
Deviation from lower limit
< 5 heart beats
6 ... 15
16 ... 30
31 ... 50
> 50 heart beats
Deviation from lower limit
< 5 heart beats
6 ... 15
16 ... 30
31 ... 50
> 50 heart beats
Speed (km/h)
0.2
0.4
0.6
0.8
1.0
Speed (km/h)
0.3
0.8
1.0
1.5
2.0
Elevation (%)
0.1
0.2
0.4
0.8
1.0
Elevation (%)
0.3
0.8
1.0
1.2
1.6
Reaction time (s)
25
25
25
20
20
Reaction time (s)
12
12
10
8
7
page 32
cos105000-2016-00 en