MODEL 630 POTENTIOMETRIC FLECTROMETER
CALIBRATION
FIGURE 11.
Block
Diagram
for
Model 630 Range Calibration.
Fully
guard
the entire
system
to
prevent
leakage
errors.
See Table
5 for
recommended equipment.
3.
50-volt
Range Calibration:
Set the Model
630 controls
as follows:
DETECTOR INPUT Switch
1010 OHMS
RANGE Switch
50
Reference
Voltage
Dials
4'9.9
9 9 10
NULL Switch
10 Mv
POLARITY Switch
+
Adjust
the Voltage
Calibrator
to apply
50 volts
dc to the Model
630.
Adjust
the 50 V
CAL potentiometer,
R334 (Figure
ZO), for
an off-null
reading
on the Model
630 equal
to
the correction
factor
at 50 volts.
4.
5-volt
Range Calibration:
set
the Model
630 controls
as follows:
DETECTOR INPUT Switch
1010 OHMS
RANGE Switch
'5
Reference
Voltage
Dials
4.9 9 9 9 10
NULL Switch
1Mv
POLARITY Switch
+
Adjust
the Voltage
Calibrator
to apply
5 volts
dc to the Model
630.
Adjust
the 5 V CAL
potentiometer,
R337 (Figure
2o),
for
an off-null
reading
on the Model
630 equal
to the
correction
factor
at 5 volts.
5.
0.5~volt
Range Calibration:
Set the Model
630 controls
as follows:
DETECTOR INPUT Switch
1010 OHMS
RANGE Switch
0.5
Reference
Voltage
Dials
.4 9 9 9 9 10
NULL Switch
0.1 MV
POLARITY Switch
+
1067~
25