Calibration; Calibration Schedule; Kelvin-Varley Divider Verification - Keithley 630 Instruction Manual

Table of Contents

Advertisement

CALIBRATION
MODEL 630 POTENTIOMETRIC ELECTROMETER
Control
Divider
Calibrate
Quad Adjustment
Meter
Calibrate
DC Balance
Oscillator
Circuit
Fig.
Refer
to
Desig.
Ref.
Paragraph
R104
15
5-4
R131
16
4-8
R155
16
4-7
R157
16
4-5
R210
18
5-7
500-volt
Calibrate
R327
20
5-5
50-volt
Calibrate
R334
20
5-5
5-volt
Calibrate
R337
20
5-5
0.5-volt
Calibrate
R339
20
5-5
TABLE 6.
Model 630 Internal
Controls.
The Table
lists
all
internal
controls,
the fig-
ure picturing
the location
and the paragraph
describing
the adjustment.
5-2.
CALIBRATION SCHEDULE.
a.
Recalibrate
the Model 630 yearly.
This
normally
means performing
the voltage
range
calibration
(paragraph
5-5).
The other
verifications
need not be done.
b.
Verify
the Kelvin-Varley
divider
(paragraph
5-3)
only
if
trouble
is suspected
in it,
or if
components
in the divider
have been replaced.
Also
make the voltage
range
calibra-
tion
after
verifying
the divider
accuracy.
c.
Check the reference
voltage
supply
stability
(paragraph
5-6)
only
if
trouble
is
suspected
in the supply
or if
some of its
components
have been replaced.
5-3.
KELVIN-VARLEY DIVIDER VERIFICATION.
a.
There
is no in-field
calibration
for
the Kelvin-Varley
divider;
its
accuracy
can
only
be verified.
The divider
accuracy
depends upon matching
resistors
and switches.
At
manufacture,
each resistor
within
the first
two Reference
Voltage
switches,
S303 and
S304, is matched
to *0.00125%.
The resistors
in the switches
are checked
as a set to an
accuracy
of better
than 0.0025%.
Individual
resistors
cannot
be replaced
without
remat-
ching
the string
at Keithley
Instruments.
b.
Kelvin-Varley
Divider
Accuracy
Verification
Procedures.
1.
Use the Model 662 Differential
Voltmeter
or equipment
with
better
limit
of
error
to match the Model 630 under
test.
Connect
the
Low end of potentiometer
R38L of
the Model 630 to the HI terminal
of the Model
662.
Connect
both
LO terminals.
NOTE
Be careful
of high
voltages
when working
within
the Model
630.
Up to 900 volts
dc is present
at various
points.
2.
Set the dials
to random settings
on both
instruments.
Settings
should
match to
iO.Ol%.
This
procedure,
however,
only
indicates
the Kelvin-Varley
divider
accuracy.
The errors
of the two instruments
may be additive,
causing
a false
verification.
The
most accurate
way is to use standard
procedures
for
checking
a Kelvin-Varley
divider
or to return
the Model 630 to Keithley
Instruments,
Inc.,
for
checking.
22
1265R

Advertisement

Table of Contents
loading

Table of Contents