CALIBRATION
MODEL 630 POTENTIOMETRIC ELECTROMETER
NOTE
The accuracy
of the Model 630 calibration
will
be no greater
than the
accuracy
of the voltage
source
used for
calibrating.
Unless
the user
is
familiar
with
techniques
for
obtaining
accuracy
greater
than 0.002%
(20
PPm) > it
is better
to return
the Model
630 to the factory
for
range
calibration.
c.
The most critical
part
in range
calibration
is establishing
a reference
source
whose accuracy
exceeds
0.002%..
Use the Model SV194B Voltage
Calibrator
for
the 500,
50,
5 and 0.5-volt
outputs.
Establish
the accuracy
of these
outputs
by determining
the cor-
rections
for
the calibrator's
CALIBRATION and OUTPUT dials
at these
outputs.
The system's
accuracy
can be determined
to approximately
10 ppm.
Added
to
the accuracy
of the standard
cell,
total
accuracy
should
be approximately
12 ppm.
Note that
this
depends upon proper-
ly executed
procedures.
d.
Procedures
to Establish
System Accuracy.
1.
Set up the system shown in Figure
10.
Use the 9.9-megohm
extender
for
only
the
0.5 and 5-volt
output.
2.
Establish
the corrections
for
the CALIBRATION dial
setting
for Model
SV194B out-
puts
of 500, 50, 5 and 0.5 volts.
Set the Model 241 Voltage
Supply
to 500 volts.
set
the Model RV722 Divider
to a voltage
equal
to that
of the standard
cell.
Adjust
the
Model
SVl94B ratio
dial
until
the null
detector
indicates
a null.
The difference
be-
tween the settings
of the Model RV722 Divider
and the ratio
dial
is the Model SVl94B
correction
factor
at 500 volts.
With this
correction,
the 500 volts
may be set to
within
approximately
12 ppm.
3.
Connect
the null
detector
to the 50, 5 and 0.5-volt
taps
of the Model
SVl94B Cal-
ibrator
in that
order.
Set the Model RV722 Divider
to corresponding
voltages.
Use the
9.9-megohm
extender
for
the 5 and 0.5-volt
outputs.
(See Figure
10).
The difference
read on the null
detector
is
the correction
factor
for
each of the three
voltages.
These voltages
may also
be set to within
12 ppm.
4.
Use the four
correction
factors
for
calibrating
the Model 630 range
setting.
f.
Procedures
for
Range Calibration.
1.
Set up the system shown in Figure
11.
The standard
cell
should
be certified
to
10 ppm.
Use the dial
correction
factors
found
for
each output.
If
the system accuracy
was properly
determined,
the input
voltage
to the Model
630 should
be correct
to 20 ppm.
2.
500-volt
Range Calibration:
Set the Model
630 controls
as follows:
DETECTOR INPUT Switch
1010 OHMS
RANGE Switch
500
Reference
Voltage
Dials
4 9 9.9 9 10
NULL Switch
VTVM
POLARITY Switch
+
Adjust
the Voltage
Calibrator
to apply
500 volts
to the Model
630.
Turn the Model 630
NULL Switch
to 10 MV and adjust
the 500 V CAL potentiometer,
R327 (Figure
ZO), for
null
on the Potentiometric
Electrometer.
24
0166R