Table of Contents USER RESPONSIBILITY ................3 INTRODUCTION ..................4 FEATURES ..................4 DESCRIPTION OF THE MANIFOLD ............5 SHIPMENT DETAILS .................5 THE COLLAPSABLE BACK BRACKET ..........5 DESCRIPTION OF PARTS .................5 COMMON TO ALL MANIFOLDS ............5 PRESSURE REGULATORS (OPERATING & LINE) ....5 PRESSURE RELIEF VALVES ..........5 PRESSURE RELIEF VALVE CHART ........6 CONTROL COMPONENTS ............6 WARNING SYSTEM COMPONENTS ........6...
Should such repair or replacement become neces- sary, please contact Amico Corporation or their distributors. All manifolds should not be repaired, or altered without prior written approval of Amico Corporation or it’s distributors. Failure to comply will void all warranty on the manifold.
E c o n o m y M a n i f o l d INTRODUCTION The Amico Economy Medical Gas Manifold incorporates the basic necessities for the distribution and monitoring of medical gases. The manifold has been designed to provide user flexibility and reliability.
The ALERT-2 economy manifold is divided into (2) main sections: COMMON TO ALL ECONOMY MANIFOLDS PRESSURE REGULATORS There are two types of regulators in the Amico manifold: the operating pressure regulator and the line pressure regulator. Both types conform to NFPA 99. . 1. Operating (Source) Regulators There are two operating regulators on every manifold, one for the left bank and one for the right bank.
System” which encompasses many types of sources of supply to nonflammable medical gas piping systems. As such, the Amico manifold is comprised of two banks of cylinders which alternately supply the pipeline, each having various control components. When the primary bank is exhausted, the pressure drops as the secondary takes over.
SAFETY FEATURES Gas Service Identification Amico manifolds are clearly labelled for the gas that they are intended to be used for. A large nameplate, indicating the appropriate gas is attached on the cabinet door. The two pipes extending from the top of the manifold, one for line pressure relief and one for the operating pressure relief, are labelled.
Page 8
When the medical gas piping system has been tested in accordance with Part 4-5 Testing of NFPA 99, the manifold can then be connected to it. The outlet pipes leading from the Amico economy manifold should be connected to their respective pipeline system connections. The connection to the relief valves should be made with a union (supplied by others) to facilitate change if required.
ARE NOT suitable for leak testing due to their inherent dangerous properties. Leak testing must be performed using either oil-free dry air or oil-free dry nitrogen. In the case of either a Medical Air or a Nitrogen Amico manifold, the actual service gas may be used to perform the leak tests as follows:...
Then turn the handle clockwise until the desired pressure is reached. NORMAL OPERATING PROCEDURE Normal operating procedure for the Amico manifold entails connecting of the proper medical gas to all cylinder extension bar outlets via the appropriate copper cylinder connection hose assemblies.
MAINTENANCE GENERAL The tests and inspections specified below apply only to the Amico manifold and not to the medical gas pipeline system as a whole. They are intended to help ensure the proper operation of the manifold and not to be interpreted as repair instructions. Fault finding and repair procedures are given in the Trouble Shooting section of this manual.
Components removed for maintenance, must be serviced, repaired and tested, only by personnel qualified to work on equipment used in medical service. Only original manufacturer’s parts, as supplied by Amico, may be used in the maintenance of Amico manifolds. S Y M P T O M...
A l e r t - 2 S e r i e s CYLINDER CHANGING PROCEDURES KEEP THE MAIN BANK VALVE OPEN THROUGHOUT THESE PROCEDURES. CLOSE CYLINDER VALVES ON ALL EMPTY CYLINDERS. DISCONNECT PIGTAILS FROM CYLINDER VALVE OUTLETS USING AN APPROPRIATE WRENCH.
E c o n o m y M a n i f o l d ORDERING INFORMATION The model number for the ALERT-2 ECONOMY MANIFOLD is as follows: M2EC-D-HH-U-GAS The letters “EC” represents that it is an Economy Manifold. The letter “D” represents the type of Line Regulator(s): D = NFPA, S = Single line. “HH”...
Page 15
A l e r t - 2 S e r i e s I T E M D E S C R I P T I O N M O D E L N U M B E R Operating check valve for M-X-MAN-33B all gases Operating pressure relief...
15.00” (381 mm) 4 holes to accommodate 4 holes to accommodate 3/8” bolts with anchors 3/8” (10 mm) bolts with anchors (Not supplied by Amico). 14.50” 14.50” (Not supplied by Amico). Hole size to be determined (368 mm) by the customer.
Page 19
14 Madison Rd., Fairfield, NJ 07004 USA Tel: 1-877-GO AMICO (1-877-462-6426) 85 Fulton Way, Richmond Hill, ON L4B 2N4 Canada Fax: (905) 764-0862 www.amico.com M2-MANUAL-ECO AUG-04...