Kanomax 4431 Instruction Manual page 55

Sound level meter
Table of Contents

Advertisement

18) Calibration signal
19) Outputs
AC output
Output
Output impedance:600Ω
Load impedance :more than 10kΩ
DC output
Output
Output impedance :50Ω
Load impedance :more than 10kΩ
20) RMS detection circuit
21) Processing
22) Pause
23) Data Storage Functions
Manual Storage :Sound level, Calculation value, Memory time,
Auto Storage
Processing Card :Storage of calculation results
24) I/O
25) Comparator Output
26) Battery Type
Battery life
Consumption current :Approx.150mA (When input 6V) at Calcutation OFF.
27) Operating temperature
28) Size
29) Weight
30) Option
・1/1 and 1/3-octave Real-time Analysis Card
Applicable standards
Measurement mode
:Electric calibration with internal oscillator
(1kHz sine wave)
:φ2.5 Jack
:1Vrms (FS)
:φ2.5 Jack
:2.5V (FS), 0.25V/10dB,
:True RMS detection circuit (computing type)
:Digital
:Normal pause function, as well as the function of
canceling the data before pausing the measurement,
are available.
:Sound pressure level or Processed values stored
in built-in Memory or Memory card
Store the Sampling Time to Built-in memory or
on Memory card
:Sampling interval 100ms, 200ms, sound level , Leq etc.
:Direct output to printer, control and output data to
computer
Digital output of real-time noise waveform with
USB interface
:Comparator Function with threshold level
:Four 1.5V Alkaline cells IEC type LR6,
Optional AC adapter
:Alkaline dry cell ; Approx.9 hours
when Switch on a back light ;
Approx.1/3
When AC adaptor is used ;Approx.2.6VA
:-10~50℃ 30%~90%RH(no condensation)
:86(W)×285(H)×46(D)
:Less than Approx.450g (Including batteries)
:JIS C 1514(IEC61260):Class1
:Sound pressure level (Lp), Equivalent continuous
Sound pressure level (Leq), Sound exposure level(L
Maximum sound pressure level (Lmax)
(One of the measurement modes selected as above is
displayed.)