4. SPECIFICATIONS

Product
Product Weight
Product Life
Safe Working Load (SWL)
Lifting Speed
Water Ingress Protection
Rating of Lift
Noise Level
Medical Electrical Equipment
Class
Protection Class
Battery
Battery Type
Battery Capacity
Battery Charging
Charger
Charger Input
Charger Output
Environmental Conditions
Operating Temperature
Battery Charging
Temperature
Operating Humidity
Storage Temperature
Storage Humidity
Atmospheric Pressure
WARNING: This equipment is not suitable in the presence of flammable
anesthetic mixtures with air or oxygen, or with nitrous oxide.
Monarch Fixed Ceiling Lift Owner's Manual
Manual trolley version: 7.5 kg (16.5 lb)
Motorized trolley version: 8.5 kg (18.7 lb)
10 years
272 kg (600 lb), 200 kg (440 lb) or 130 kg (286 lb)
5.5 cm/sec (2.2 in/sec) at 0 kg
5 cm/sec (2 in/sec) at 130 kg (286 lb)
4.5 cm/sec (1.8 in/sec) at 200 kg (440 lb)
4.0 cm/sec (1.6 in/sec) at 272 kg (600 lb)
IP21
Maximum 54 dBA
Class 1
Type BF
Lithium ion, 25.2V 2500 mAh
Approximately 30 cycles of 61 cm (24 in) at 272 kg
(600 lb)
Approximately 45 cycles of 61 cm (24 in) at 200 kg
(440 lb)
Approximately 70 cycles of 61 cm (24 in) at 130 kg
(286 lb)
Full capacity in approximately 2 hours
100-240 VAC, 50-60Hz
29.4 VDC, max 1A
5 °C to 40 °C (41 °F to 104 °F)
10 °C to 40 °C (50 °F to 104 °F)
15 to 93%, non-condensing
-25 °C to 70 °C (-13 °F to 158 °F)
0 to 93%, non-condensing
70 kPa to 106 kPA
Part No. 001108-EN, 21-m09-2017