Reference Manual 00809-0200-4802, Rev BA May 2013 Rosemount 3051S Wireless Series Pressure, Level, and Flow Solutions with IEC 62951 (WirelessHART™) Protocol...
Page 3
The products described in this document are NOT designed for nuclear-qualified applications. Using non-nuclear qualified products in applications that require nuclear-qualified hardware or products may cause inaccurate readings. For information on Rosemount nuclear-qualified products, contact your local Emerson Process Management Sales Representative.
Page 4
cause electrical shock. The Rosemount 3051S and all other wireless devices should be installed only after the Smart Wireless Gateway has been installed and is functioning properly. Wireless devices should also be powered up in order of proximity from the Smart Wireless Gateway, beginning with the closest.
Reference Manual Table of Contents 00809-0200-4802, Rev BA May 2013 Table of contents 1Section 1: Introduction Using This Manual ............1 Models Covered .
Page 6
Installing the LCD display .........40 Rosemount 304, 305 and 306 Integral Manifolds ......41 3.5.1...
Page 7
Reference Manual Table of Contents 00809-0200-4802, Rev BA May 2013 5Section 5: Operation and Maintenance Overview .............51 Calibration .
Page 8
Reference Manual Table of Contents 00809-0200-4802, Rev BA May 2013 B.1.1 Approved manufacturing locations....... 139 B.1.2 Telecommunication compliance .
Product Recycling/Disposal ..........page 4 Using This Manual The sections in this manual provide information on installing, operating, and maintaining the ™ Rosemount 3051S Wireless pressure transmitter with IEC 62591 (WirelessHART ) protocol. The sections are organized as follows: Section 2: Configuration provides instruction on commissioning and operating 3051S ...
Reference Manual Section 1: Introduction 00809-0200-4802, Rev BA May 2013 Models Covered The following Rosemount 3051S Wireless pressure transmitters are covered in this manual. ™ Rosemount 3051S Wireless Coplanar Pressure Transmitter Measurement Type Performance Class Differential Gage Absolute Classic Ultra Ultra for Flow –...
Reference Manual Section 1: Introduction 00809-0200-4802, Rev BA May 2013 Service support To expedite the return process outside of the United States, contact the nearest Emerson Process Management representative. Within the United States, call the Emerson Process Management Instrument and Valves Response Center using the 1-800-654-RSMT (7768) toll-free number.
Reference Manual Section 1: Introduction 00809-0200-4802, Rev BA May 2013 Product Recycling/Disposal Recycling of equipment and packaging should be taken into consideration and disposed of in accordance with local and national legislation/regulations. Introduction...
Verify the latest Device Driver (DD/DTM) is loaded on your systems to ensure proper communications. Download the latest DD at www.emersonprocess.com. 2. In the Browse by Member dropdown menu, select Rosemount business unit of Emerson Process Management. 3. Select desired Product Within...
Figure 2-1 on page Bench top configuration consists of testing the transmitter and verifying transmitter configuration data. 3051S Wireless transmitters must be configured before installation. Configuring the transmitter on the bench before installation using a 375 Field Communicator, AMS, or any IEC 62591 (WirelessHART) Communicator ensures that all network settings are working correctly.
P/N 00753-9200-0010 P/N 00753-9200-0010 P/N 00753-9200-0010 COMM COMM COMM For HART communication, a 3051S WirelessHART DD is required. Device network configuration 2.5.1 Join device to network Fast Keys 2, 1, 3 In order to communicate with the Smart Wireless Gateway, and ultimately the Host System, the transmitter must be configured to communicate over the wireless network.
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 2.5.2 Configure update rate Fast Keys 2, 1, 4 The Update Rate is the frequency at which a new measurement is taken and transmitted over the wireless network. This by default is 1 minute. This may be changed at commissioning, or at any time via AMS Wireless Configurator.
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 Review configuration data The following is a list of factory default configurations that can be viewed by using the Field Communicator or AMS. Follow the steps below to review the transmitter configuration information.
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 2.6.2 Review Device and Sensor Information Fast Keys 2, 2, 9 To view device information: From the Home screen, choose 2: Configure. Choose 2: Manual Setup. Choose 9: Device Information. Choose from the corresponding number to view each field: Identification Revisions...
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 2.6.3 Review radio information Fast Keys 2, 2, 9, 3 To view radio information: From the Home screen, choose 2: Configure. Choose 2: Manual Setup. Choose 9: Device Information. Choose 3: Radio. Select from the corresponding number to view each field: Manufacturer Device Type...
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 The Operating Parameters menu displays the following information pertaining to the device: Percent of Range Pressure Pressure Status Sensor Temperature Sensor Temperature Status Electronics Temperature ...
The Rosemount 3051S Wireless has two output settings: Linear and Square Root. Activate the square root output option to make output proportional to flow. As input approaches zero, the 3051S Wireless automatically switches to linear output in order to ensure a more smooth, stable output near zero (see Figure 2-2).
In the 3051S Wireless pressure transmitter, damping only takes effect when the device is placed in high power refresh mode and during calibration. In normal power mode, the effective damping is 0.
The LCD display indicates output and abbreviated diagnostic messages. Note Use Rosemount Wireless LCD Part Number: 00753-9004-0002. The LCD display features a four-line display and a bar graph. The first line of five characters displays the output description, the second line of seven digits displays the actual value, the third line of six characters displays engineering units and the fourth line displays “Error”...
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 2.10 Detailed setup 2.10.1 Configure process alarms Fast Keys 2, 1, 6 Alerts allow the user to configure the transmitter to output a HART message when the configured data point is exceeded. A process alert will be transmitted continuously if the set points are exceeded and the alert mode is ON.
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 2.10.2 Sensor temperature unit Fast Keys 2, 2, 8, 3 The Sensor Temperature Unit command selects between Celsius and Fahrenheit units for the sensor temperature. The sensor temperature output is accessible via HART only. To select the sensor temperature unit: From the Home screen, choose 2: Configure.
Page 27
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 Note If Scaled Variable is mapped as the primary variable and square root mode is selected, ensure transfer function is set to linear. Refer to “Set transfer function” on page Field Communicator v3.3 To configure the scaled variable with a Field Communicator, perform the following procedure: From the HOME screen follow the fast key sequence “Scaled Variable Configuration.”...
Page 28
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 Enter linear offset value in scaled variable (custom) units (If performing a Linear Function). Skip to step 11. Enter Low Flow cutoff mode (If performing a Square Root Function) Choose OFF if a low flow cutoff value is not desired. b.
Page 29
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 DP Level example Figure 2-3. Example tank 20 mA 230 in. 0.94 sg 200 in. 4 mA 12 in. A differential transmitter is used in a level application where the span is 188 inH O (200 in.
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 value is 1000 gallons of water per hour. Based on this information, the Scaled Variable configuration would be as follows: Scaled Variable units: gal/h Scaled data options: square root Pressure value position 125 inH2O (311 mbar) Scaled Variable position 20,000 gal/h (75,708...
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 Re-mapping using a Field Communicator From the HOME screen, enter the fast key sequence 2, 1, 7 Fast Keys: guided set up 2, 2, 3, 5 Fast Keys: manual set up Re-mapping using AMS Device Manager Right click on the device and select Configure.
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 2.11.2 Join status Fast Keys 3, 4, 1 To view the join status of the device, perform the following procedure: From the Home screen, choose 3: Service Tools. Choose 4: Communications. Choose 1: Join Status.
Use the cloning feature of the Field Communicator or the AMS “User Configuration” feature to configure several 3051S Wireless transmitters similarly. Cloning involves configuring a transmitter, saving the configuration data, then sending a copy of the data to a separate transmitter.
Page 34
Reference Manual Section 2: Configuration 00809-0200-4802, Rev BA May 2013 When finished, the Field Communicator informs you of the status. Repeat Steps through configure another transmitter. Note The transmitter receiving cloned data must have the same software version (or later) as the original transmitter.
Installing the LCD display ..........page 40 Rosemount 304, 305 and 306 Integral Manifolds ......page 41 Overview The information in this section covers installation considerations.
Do not attempt to loosen or remove process connectors while the transmitter is in service. Replacement equipment or spare parts not approved by Rosemount Inc. for use as spare parts could reduce the pressure retaining capabilities of the transmitter and may render the instrument dangerous.
Reference Manual Section 3: Installation 00809-0200-4802, Rev BA May 2013 Considerations 3.3.1 General Measurement performance depends upon proper installation of the transmitter and impulse piping. Mount the transmitter close to the process and use a minimum of piping to achieve best performance.
Figure 3-2. Conduit Plug Location Conduit Plug Conduit Plug Field Communicator connections In order for the Field Communicator to interface with the Rosemount 3051S, the Power Module must be connected. Figure 3-3. 375 Field Communicator Connections P/N 00753-9200-0010 P/N 00753-9200-0010...
Page 39
Reference Manual Section 3: Installation 00809-0200-4802, Rev BA May 2013 Figure 3-4. Installation Flowchart START HERE Bench Calibration? Field Install Check Switches page 39 Configure Verify Mount Transmitter Review Install Power Set Units Transmitter Module Configuration (pages – Check Process Set Range Connection Points...
Cover installation Always ensure a proper seal by installing the electronics housing cover(s) so that metal contacts metal. Use Rosemount O-rings. Conduit plugs Replace temporary orange plugs with included conduit plugs (found in the box) using approved thread sealant.
Page 41
Position the antenna such that it is vertical, typically straight up (antenna may be pointed down as well.) Flange bolts The 3051S can be shipped with a Coplanar flange or a Traditional flange installed with four 1.75-inch flange bolts. Mounting bolts and bolting configurations for the Coplanar and Traditional flanges can be found on page 33.
Page 42
Reference Manual Section 3: Installation 00809-0200-4802, Rev BA May 2013 GAGE/ABSOLUTE TRANSMITTER DIFFERENTIAL TRANSMITTER Drain/Vent Plug Drain/Vent Drain/Vent 1.75 (44) x 4 1.75 (44) x 4 1.50 (38) x 4 1.50 (38) x 4 Note Dimensions are in inches (millimeters). Installation...
Page 43
Flange Bolts 1.75 (44) Adapter Bolts 1.50 (38) Flange/ Adapter Bolts 2.88(73) (1) DIN-compliant traditional flange requires 1.75 in. (44 mm) length adapter bolts. (2) Rosemount 3051S In-line transmitters are direct mount and do not require bolts for process connection. Installation...
Page 44
Reference Manual Section 3: Installation 00809-0200-4802, Rev BA May 2013 Impulse piping The piping between the process and the transmitter must accurately transfer the pressure to obtain accurate measurements. There are five possible sources of error: leaks, friction loss (particularly if purging is used), trapped gas in a liquid line, liquid in a gas line, and density variations between the legs.
Page 45
Reference Manual Section 3: Installation 00809-0200-4802, Rev BA May 2013 Mounting requirements Refer to Figure 3-5 for examples of the following mounting configurations: Liquid flow measurement Place taps to the side of the line to prevent sediment deposits on the process isolators. ...
The two flange adapters are distinguished by unique o-ring grooves. Only use the o-ring that is designed for its specific flange adapter, as shown below. ROSEMOUNT 3051S / 3051 / 2051 / 3001 / 3095 / 2024 Flange Adapter...
Reference Manual Section 3: Installation 00809-0200-4802, Rev BA May 2013 3.4.3 Consider housing rotation The housing can be rotated to improve field access to wiring or to better view the optional LCD display. Perform the following procedure: Figure 3-6. Housings ®...
Installing the LCD display Transmitters ordered with the LCD display will be shipped with the display installed. Note Only use Rosemount Wireless LCD Part Number: 00753-9004-0002 Note An LCD from a wired device will not function in a wireless device.
The Rosemount 306 Integral Manifold is used with In-line transmitters to provide block-and-bleed valve capabilities of up to 10000 psi (690 bar). The Rosemount 304 comes in two basic styles: traditional (flange x flange and flange x pipe) and wafer. The 304 traditional manifold comes in 2, 3, and 5-valve configurations.
Maintenance, “Zero Trim” on page 3.5.2 Rosemount 306 Integral Manifold installation procedure The 306 Manifold is for use only with a 3051S Wireless In-line transmitter. Assemble the 306 Manifold to the 3051S Wireless In-line transmitter with a thread sealant. Installation...
3.5.3 Rosemount 304 Conventional Manifold installation procedure To install a Rosemount 304 Conventional Manifold to a Rosemount 3051S Wireless transmitter: Align the Conventional Manifold with the transmitter flange. Use the four manifold bolts for alignment. Finger tighten the bolts, then tighten the bolts incrementally in a cross pattern to final torque value.
Page 52
Vent equalizing valve will be closed. Valve Valve Equalize (closed) Isolate Isolate (open) (open) Process To zero the 3051S, close the block valve to the low Drain/ Drain/ pressure (downstream) side Vent Vent of the transmitter first. Valve Valve Equalize...
Page 53
Equalize Equalize (closed) (closed) Isolate Isolate (open) (open) Process Process Drain Vent (closed) To zero the 3051S, first close the block valve on the low pressure (downstream) side Test Test of the transmitter. (Plugged) (Plugged) Equalize Equalize (closed) (closed) Isolate...
Page 54
Reference Manual Section 3: Installation 00809-0200-4802, Rev BA May 2013 Open the equalize valve on the low pressure (downstream) side of the Test Test transmitter. The manifold is (Plugged) (Plugged) now in the proper Equalize Equalize configuration for zeroing the (open) (open) transmitter.
Note The Rosemount 3051S Wireless and all other wireless devices should be installed only after the Smart Wireless Gateway has been installed and is functioning properly. Wireless devices should also be powered up in order of proximity from the Smart Wireless Gateway, beginning with the closest device to the Smart Wireless GatewaySmart.
May 2013 Network status If the Rosemount 3051S Wireless was configured with the Network ID and Join Key and sufficient time for network polling has passed, the transmitter should be connected to the network. To verify connectivity, open the Smart Wireless Gateway’s integral web interface and navigate to the Explorer page.
Page 57
N E G O T L I M - O P Field communicator To verify device operation using a HART Field Communicator, a 3051S Wireless DD is required. For connecting with a 375 Field Communicator, refer to Figure 2-1 on page Table 4-1.
Reference Manual Section 4: Commissioning 00809-0200-4802, Rev BA May 2013 A green status indicator means that the device is working properly. A red indicator means that there is a problem with either the device or its communication path. For more detail on a specific device, click on the tag name.
The 3051S SuperModule uses a microprocessor that contains information about the sensor’s specific characteristics in response to pressure and temperature inputs. A smart transmitter compensates for these sensor variations. The process of generating the sensor performance profile is called factory sensor characterization.
Note Do not perform a zero trim on 3051S Wireless absolute pressure transmitters. Zero trim is zero based, and absolute pressure transmitters reference absolute zero. To correct mounting position effects on a 3051S Wireless absolute pressure transmitter, perform a low trim within the sensor trim function.
The trim values allow you to optimize performance over your specified measuring range at the calibration temperature. During a trim operation, the 3051S Wireless is placed in high power refresh mode, which provides frequent pressure measurement updates and allows the configured damping to take effect.
0, 1, 2, and 3) do not require this procedure because optimization occurs in the sensor. Applying high static pressure to the 3051S Wireless Range 4 and Range 5 pressure transmitters causes a systematic shift in the output. This shift is linear with static pressure; correct it by performing the “Sensor trim”...
Page 63
Reference Manual Section 5: Operation and Maintenance 00809-0200-4802, Rev BA May 2013 For line pressures above 2000 psi (138 bar), the zero effect error is ± 0.2% of the upper range limit plus an additional ± 0.2% of upper range limit error for each 1000 psi (69 bar) of line pressure above 2000 psi (138 bar).
Page 64
HT = 1518 inH O (3.78 bar) Complete a Rosemount 3051S Wireless sensor trim and enter the corrected values for low trim (LT) and high trim (HT), refer to “Sensor trim” on page Enter the corrected input values for low trim and high trim through the Field Communicator keypad after you apply the nominal value of pressure as the transmitter input.
May 2013 LCD screen messages 5.3.1 Startup screen sequence The following screens will display when the Power Module is first connected to the Rosemount 3051S Wireless. X X X X X All Segments On: used to visually determine if there...
Page 66
Reference Manual Section 5: Operation and Maintenance 00809-0200-4802, Rev BA May 2013 S N S R SV Screen - sensor temperature value 2 5. 0 0 d e g c D E V TV Screen - device temperature value 2 5. 2 5 d e g c S u p l y QV Screen - voltage reading at the power supply...
Reference Manual Section 5: Operation and Maintenance 00809-0200-4802, Rev BA May 2013 5.3.2 Diagnostic Button Screen Sequence The following five screens will display when the device is operating properly and the Diagnostic Button has been pressed. Device Information - Tag: user entered tag which is A b c d e eight characters long - will not display if all characters are blank...
Reference Manual Section 5: Operation and Maintenance 00809-0200-4802, Rev BA May 2013 S u p l y Diagnostic Button Screen 5: voltage reading at the 7. 2 1 power supply terminals v o l t s 5.3.3 Network diagnostic status screens These screens display the network status of the device.
Page 69
Reference Manual Section 5: Operation and Maintenance 00809-0200-4802, Rev BA May 2013 Diagnostic Button Screen 4.4: the device is in a n e t w k disconnected state and requires a “Force Join” command to join the network D I S C N T Diagnostic Button Screen 4.5: the device is N E T w K searching for the Network...
Page 70
Reference Manual Section 5: Operation and Maintenance 00809-0200-4802, Rev BA May 2013 Diagnostic Button Screen 4.9: the device has n e t w k joined a network and has been fully configured and has multiple parents Operation and Maintenance...
Reference Manual Section 5: Operation and Maintenance 00809-0200-4802, Rev BA May 2013 5.3.4 Device diagnostic screens The following screens will show the device diagnostics depending on the state of the device. Device Information - Status: there is a critical error D E V which may prevent the device from operating correctly.
Page 72
Reference Manual Section 5: Operation and Maintenance 00809-0200-4802, Rev BA May 2013 S u p l y QV Screen - voltage reading at the power supply 7. 2 1 terminals v o l t s P R C N T Percent Range Screen - percent range reading 7.
Page 73
Reference Manual Section 5: Operation and Maintenance 00809-0200-4802, Rev BA May 2013 Diagnostic Button Screen 7.1: the terminal voltage has dropped below level of operating limit. Replace s u p l y the Power Module (Part Number: 00753-9220-0001) f a i l u r Diagnostic Button Screen 7.2: the terminal voltage is below the recommended operating range - if this is s u p l y...
Page 74
- s n s r check the process, and sensor wiring connections - check additional status for more detailed information of the warning source w a r n Note Use the Rosemount Wireless LCD Part Number: 00753-9004-0002. Operation and Maintenance...
Install and tighten process connectors before applying pressure. Do not attempt to loosen or remove process connectors while the transmitter is in service. Table 6-1. Rosemount 3051S Wireless Device Status Information Device Status Description Recommended Action Reset the device...
Page 77
Reference Manual Section 6: Troubleshooting 00809-0200-4802, Rev BA May 2013 Device Status Description Recommended Action Check process for possible saturation condition Pressure has The sensor has exceeded the Exceeded Limits maximum measurement range. Verify that the appropriate sensor was chosen for the application Reconfirm sensor configuration Reset the device...
Page 78
Reference Manual Section 6: Troubleshooting 00809-0200-4802, Rev BA May 2013 Device Status Description Recommended Action Verify that the process variable is within user LO Alarm The primary variable has surpassed the user defined limit. specified limits Reconfirm the user defined alarm limit If not needed, disable this alert Verify that the process variable is within user LO LO Alarm...
Page 79
Reseat the LCD according to Verify that the LCD display is a wireless LCD Meter. An LCD from a wired device will not function in a wireless device. Rosemount part number: 00753-9004-0002 Verify that the LCD display mode is not disabled.
Leave the process connection in place and ready for re-installation. The Rosemount 3051S Wireless In-line transmitter is attached to the process by a single hex nut process connection. Loosen the hex nut to separate the transmitter from the process.
Reference Manual Section 6: Troubleshooting 00809-0200-4802, Rev BA May 2013 6.3.3 Removing the feature assembly and the SuperModule from the housing The Wireless Feature Board is located in the electronics compartment. To remove the assembly, perform the following procedure. Remove the housing cover opposite the field terminal side. Remove the LCD Display, if applicable.
Reference Manual Section 6: Troubleshooting 00809-0200-4802, Rev BA May 2013 Important To prevent damage to the SuperModule cable, disconnect it from the PlantWeb assembly before you remove the SuperModule from the housing. Loosen the housing rotation set screw with a -inch hex wrench, then rotate back one full turn.
Reference Manual Section 6: Troubleshooting 00809-0200-4802, Rev BA May 2013 6.4.3 Reassemble the process flange Inspect the Teflon SuperModule o-rings. If the o-rings are not damaged, reuse them. Replace o-rings that show any signs of damage, such as nicks, cuts, or general wear. Note If you are replacing the o-rings, be careful not to scratch the o-ring grooves or the surface of the isolating diaphragm when removing the damaged o-rings.
Page 84
Reference Manual Section 6: Troubleshooting 00809-0200-4802, Rev BA May 2013 Table 6-4. Bolt Installation Torque Values Bolt Material Initial Torque Value Final Torque Value CS-ASTM-A445 Standard 300 in-lb. (34 N-m) 650 in-lb. (73 N-m) 316 SST—Option L4 150 in-lb. (17 N-m) 300 in-lb.
Appendix A: Specifications and reference data Reference Manual May 2013 00809-0200-4802, Rev BA Appendix A Specifications and Reference Data Performance specifications ..........page 77 Functional specifications .
Page 86
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Reference accuracy (1)(2) Models Ultra Classic Ultra for Flow 3051S_CD, CG Ranges 2 - 4 ±0.025% of span. ±0.055% of span. ±0.04% of reading up to 8:1 For spans less than 10:1, For spans less than 10:1, DP turndown from URL;...
Page 87
(1) Warranty details can be found in Emerson Process Management Terms & Conditions of Sale, Document 63445, Rev G (10/06). (2) Rosemount Ultra and Ultra for Flow transmitters have a limited warranty of twelve (12) years from date of shipment. All other provisions of Emerson Process Management standard limited warranty remain the same.
Page 88
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Ambient temperature effect Models Ultra Classic Ultra for Flow 3051S_CD, CG per 50 °F (28 °C) per 50 °F (28 °C) per 50 °F (28 °C) Range 2 - 5 ±...
Page 89
Range 0 Range 1 (1) For line pressures above 2000 psi (137,9 bar) and 3051S_CD Ranges 4-5, see the 3051S Reference Manual (document number 00809-0100-4801). (2) Zero error can be calibrated out. (3) Specifications for option code P0 are 2 times those shown above.
Appendix A: Specifications and reference data Reference Manual May 2013 00809-0200-4802, Rev BA A.1.2 Functional specifications Range and sensor limits 3051S_ Minimum Span 3051S_ Range and Sensor Limits Ultra and Classic Upper (URL) Lower (LRL) Ultra for (3)(4) 3051S_CD 3051S_CG, LG 3051S_LD Flow 0.1 inH...
Page 91
-40 °F or 185 °F (-40 °C or 85 °C) may reduce specified life by less than 20 percent. Overpressure Limits Transmitters withstand the following limits without damage: Rosemount 3051S_CD, CG Range 0: 750 psi (51,7 bar) Range 1: 2000 psig (137,9 bar) Ranges 2–5: 3626 psig (250,0 bar)
Page 92
At 248 °F (120 °C), the rating decreases with increasing temperature, per DIN 2401. Static pressure limit Rosemount 3051S_CD Only Operates within specifications between static line pressures of 0.5 psia and 3626 psig; 4500 psig (310,3 bar) for option code P9 6092 psig (420 bar) for option code P0 (3051S2CD only) Range 0: 0.5 psia to 750 psig (0,03 to 51,71 bar)
Page 93
–40 to 300 °F (–40 to 149 °C) (4) (5) Inert Fill Sensor -40 to 185 °F (–40 to 85 °C) Rosemount 3051S_T In-Line (Process Fill Fluid) Silicone Fill Sensor –40 to 250 °F (–40 to 121 °C) Inert Fill Sensor –22 to 250 °F (–30 to 121 °C)
Page 94
In the 3051S Wireless pressure transmitter, damping only takes effect when the device is placed in high power refresh mode and during calibration. In normal power mode, the effective damping is 0.
Electrical connections –14 NPT, G , and M20 x 1.5 (CM20) conduit. HART interface connections fixed to terminal block. Process connections Rosemount 3051S_C –18 NPT on 2 -in. centers –14 NPT and RC on 2-in. (50.8 mm), 2 -in. (54.0 mm), or 2 -in.
Page 96
00809-0200-4802, Rev BA May 2013 Wetted O-rings Glass-filled PTFE (Graphite-filled PTFE with Isolating Diaphragm code 6) Rosemount 3051S_L process wetted parts Flanged Process Connection (Transmitter High Side) Process diaphragms, including process gasket surface 316L SST, Alloy C-276, or Tantalum Extension CF-3M (Cast 316L SST, material per ASTM A743), or CW-12MW (Cast C-276, material ASTM A494);...
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Sensor module fill fluid Silicone or inert halocarbon (Inert is not available with 3051S_CA). In-Line series uses ® Fluorinert FC-43. Process fill fluid (liquid level only) 3051S_L: Syltherm XLT, D.C. Silicone 704, D.C.
Page 98
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Shipping weights for Rosemount 3051S Wireless Pressure Transmitter Table A-1. SuperModule Platform weights SuperModule Platform Weight in lb. (kg) Coplanar 3.1 (1,4) In-Line 1.4 (0,6) (1) Flange and bolts not included.
Page 99
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Option Code Option Add lb (kg) Level Flange—2 in., 300 8.2 (3,7) DIN Level Flange, SST, DN 50, PN 40 7.8 (3,5) DIN Level Flange, SST, DN 80, PN 40 13.0 (5,9) (1) Includes LCD display and display cover.
00809-0200-4802, Rev BA May 2013 Dimensional drawings Dimensions are in inches (millimeters). Process adapters (option D2) and Rosemount 305 integral manifolds must be ordered with the transmitter. Wireless PlantWeb Housing with In-Line and Coplanar SuperModule Platform Extended Range, External Antenna 90°...
Page 101
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Table A-5. 3051S_L Dimensional Specifications Except where indicated, dimensions are in inches (millimeters). Class Pipe Flange Bolt Circle Outside No. of Bolt Hole Extension O.D. Size Thickness A Diameter Diameter Bolts...
00809-0200-4802, Rev BA May 2013 Ordering Information Table A-6. Rosemount 3051S Coplanar Differential, Gage, or Absolute Transmitter ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 103
Standard ★ None ★ Assemble to Rosemount 305 Integral Manifold ★ Assemble to Rosemount 304 or AMF Manifold and SST traditional flange (7)(8)(9) ★ Assemble to one Rosemount 1199 Diaphragm Seal (7)(8)(9) ★ Assemble to two Rosemount 1199 Diaphragm Seals ★...
Page 104
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 ★ Traditional flange 316 SST ★ Traditional flange Cast C-276 Alloy C-276 ★ Traditional flange Cast Alloy Alloy 400/K-500 ★ Traditional flange Alloy C-276 ★ DIN-compliant traditional –18 NPT 316 SST -in.
Page 105
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 ★ Junction Box housing –14 NPT ★ Junction Box Housing with output for remote display and interface Aluminum –14 NPT ★ Junction Box Housing with output for remote display and interface Aluminum M20 x 1.5 ★...
Page 106
Custom alarm and saturation signal levels, high alarm (Requires C1 and Configuration Data Sheet) (17) ★ Custom alarm and saturation signal levels, low alarm (Requires C1 and Configuration Data Sheet) (17) ★ Low alarm (standard Rosemount alarm and saturation levels) Hardware Adjustments Standard Standard (17)(18) ★...
Page 107
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Custody Transfer Standard Standard (19) ★ Measurement Canada Accuracy Approval External Ground Screw Assembly Standard Standard ★ External ground screw assembly Drain/Vent Valve Standard Standard (16) ★ Delete transmitter drain/vent valves (install plugs) Expanded (16)
Page 108
Cleaning for special services (16) Cleaning for less than 1PPM chlorine/fluorine Bolting Material Standard Standard ★ 4500 psig (310 bar) static pressure limit (Rosemount 3051S_CD only) ★ (29) 6092 psig (420 bar) static pressure limit (Rosemount 3051S2CD only) Specifications and Reference Data...
Page 109
(13) Long-Life Power Module must be shipped separately, order Part #00753-9220-0001 or order 701PBKKF. (14) Requires PlantWeb housing and output code A. Includes Hardware Adjustments as standard. (15) Requires Rosemount Engineering Assistant to configure. (16) Not available with process connection option code A11.
Page 110
(28) P1 is not available with 3051S_CA0. (29) Requires 316L SST, Alloy C-276, or Gold-plated 316L SST diaphragm material, assemble to Rosemount 305 integral manifold or DIN-compliant traditional flange process connection, and bolting option L8. Limited to Pressure Range (Differential), ranges 2A – 5A.
Page 111
Material Vent Standard Standard ★ Assemble to Rosemount 306 Integral Manifold (7)(5)(6) ★ Assemble to one Rosemount 1199 Diaphragm Seal ★ –14 NPT female –18 NPT ★ A DIN 16288 male (Range 1-4 only) 2-in. ANSI class 150 Expanded Traditional flange,...
Page 112
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Expanded PlantWeb housing Aluminum PlantWeb housing Junction Box housing Aluminum Junction Box Housing with output for remote display and Aluminum interface Wireless Options (Requires option code X and wireless PlantWeb housing) Update Rate Standard Standard...
Page 113
Custom alarm and saturation signal levels, high alarm (Requires C1 and Configuration Data Sheet) (17) ★ Custom alarm and saturation signal levels, low alarm (Requires C1 and Configuration Data Sheet) (17) ★ Low alarm (standard Rosemount alarm and saturation levels) Hardware Adjustments Standard Standard (17)(13) ★...
Page 114
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 ★ IECEx Flameproof, Dust Ignition-proof ★ IECEx Intrinsic Safety ★ IECEx FISCO Intrinsic Safety (F fieldbus protocol only) OUNDATION ★ IECEx Type n ★ IECEx Flameproof, Dust Ignition-proof, Intrinsic Safety, Type n ★...
Page 115
(19) Not available with Housing code 01 or 7J. (20) Not available with output code F, Housing code 01, option code DA1, or option code QT. (21) See the 3051S Reference Manual (document number 00809-0100-4801) for cable requirements. Contact an Emerson Process Management representative for additional information.
Page 116
Reference Manual May 2013 00809-0200-4802, Rev BA Table A-8. Rosemount 3051SFA Annubar Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 117
Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Table A-8. Rosemount 3051SFA Annubar Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 118
Reference Manual May 2013 00809-0200-4802, Rev BA Table A-8. Rosemount 3051SFA Annubar Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 119
Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Table A-8. Rosemount 3051SFA Annubar Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 120
★ Classic MV: 0.85% flow rate accuracy, 8:1 flow turndown, 5-yr. stability 3051S Single Variable SuperModule, Measurement Types 3, 4, 7, and D (14) ★ Ultra: up to 0.9% flow rate accuracy, 8:1 flow turndown, 10-year stability, limited 12-year warranty ★...
Page 121
European Pressure Directive (PED) Installed in Flanged Pipe Spool Section Expanded 150# Flanged Connection with Rosemount Standard Length and Schedule 300# Flanged Connection with Rosemount Standard Length and Schedule 600# Flanged Connection with Rosemount Standard Length and Schedule Instrument Connections for Remote Mount Option...
Page 122
Reference Manual May 2013 00809-0200-4802, Rev BA Table A-8. Rosemount 3051SFA Annubar Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 123
(20) ★ Custom Alarm and Saturation Signal Levels, Low Alarm Note: A Configuration Data Sheet (00806-0100-4809) must be completed. (20) ★ Low Alarm (Standard Rosemount Alarm and Saturation Signal Levels) Special Configuration (Hardware) Standard Standard (20)(26) ★ Hardware Adjustment (zero, span, security) ★...
Page 124
(option code I7). Contact an Emerson Process Management representative for additional information. (13) Available with output code A only. Not available with approvals. See Rosemount 753R Product Data Sheet, 00813-0100-4379, to specify Web-Based Monitoring Indicator. Does not integrate into plant host systems.
Page 125
Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Table A-9. Rosemount 3051SFC Compact Orifice Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 126
Expanded (9)(12) Assemble to Rosemount 753R Web-based Monitoring Indicator Transmitter Performance Class 3051S MultiVariable SuperModule, Measurement Types 1, 2, 5, and 6 Standard Standard ★ Ultra for Flow: 0.55% flow rate accuracy, 14:1 flow turndown, 10-yr stability. limited 12-yr warranty ★...
Page 127
The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time. Transmitter Performance Class 3051S Single Variable SuperModule, Measurement Types 3, 4, 7, and D Standard Standard (13) ★...
Page 128
Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Table A-9. Rosemount 3051SFC Compact Orifice Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 129
Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Table A-9. Rosemount 3051SFC Compact Orifice Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 130
(option code I7). Contact an Emerson Process Management representative for additional information. (12) Available with output code A only. Not available with approvals. See Rosemount 753R Product Data Sheet, 00813-0100-4379, to specify Web-Based Monitoring Indicator. Does not integrate into plant host systems.
Page 131
Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Table A-10. Rosemount 3051SFP Integral Orifice Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 132
Reference Manual May 2013 00809-0200-4802, Rev BA Table A-10. Rosemount 3051SFP Integral Orifice Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 133
Ultra for Flow: 0.80% flow rate accuracy, 14:1 flow turndown, 10-year stability. limited 12-year warranty ★ Classic 2: 0.85% flow rate accuracy, 8:1 flow turndown, 5-year stability Transmitter Performance Class 3051S Single Variable SuperModule, Measurement Types 3, 4, 7, and D Standard Standard (11) ★...
Page 134
Reference Manual May 2013 00809-0200-4802, Rev BA Table A-10. Rosemount 3051SFP Integral Orifice Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 135
Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Table A-10. Rosemount 3051SFP Integral Orifice Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 136
Reference Manual May 2013 00809-0200-4802, Rev BA Table A-10. Rosemount 3051SFP Integral Orifice Flowmeter Ordering Information ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 137
(8) Material specified is cast as follows: CF-8M is the cast version of 316 SST, CF-3M is the cast version of 316L SST. For housing, material is aluminum with polyurethane paint. (9) Available with output code A only. Not available with approvals. See Rosemount 753R Product Data Sheet, 00813-0100-4379, to specify Web-Based Monitoring Indicator.
Page 138
Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Table A-11. Rosemount 3051S Liquid Level Transmitter ★ The Standard offering represents the most common models and options. These options should be selected for best delivery. __The Expanded offering is subject to additional delivery lead time.
Page 139
Standard Standard ★ Tuned-System Assembly, One Capillary Remote Seal (Requires 1199 model number, see Table 6 of Rosemount DP Level PDS for seal information) ★ 316L SST isolator / SST transmitter flange ★ Alloy C-276 isolator / SST transmitter flange...
Page 140
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Seal Fill Fluid (High Side) Temperature Limits (Ambient Temperature of 70° F (21° Standard Standard ★ Syltherm XLT -102 to 293 °F (-75 to 145 °C) ★ Silicone 704 32 to 401 °F (0 to 205 °C) ★...
Page 141
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Lower Housing Material for FF, Extension Length for EF (High Side) Flush Flanged Seal Extended Flanged Seal Standard Standard ★ None — ★ — 2-in. (50 mm) ★...
Page 142
Custom alarm and saturation signal levels, high alarm (Requires C1 and Configuration Data Sheet) (11) ★ Custom alarm and saturation signal levels, low alarm (Requires C1 and Configuration Data Sheet) ★ (11) Low alarm (standard Rosemount alarm and saturation levels) Specifications and Reference Data...
Page 143
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 Hardware Adjustments Standard Standard (11)(12) ★ Hardware adjustments (zero, span, alarm, security) Flange Adapter Standard Standard ★ -14 NPT flange adapter Expanded SST flange adapter Custody Transfer Standard Standard (13)
Page 144
Reference Manual Appendix A: Specifications and reference data 00809-0200-4802, Rev BA May 2013 ★ CSA FISCO Intrinsically Safe (F fieldbus protocol only) OUNDATION ★ IECEx FISCO Intrinsic Safety (F fieldbus protocol only) OUNDATION ★ ATEX Flameproof, Intrinsic Safety, Type n, Dust ★...
Page 145
(option code I7). Contact an Emerson Process Management representative for additional information. (4) With option code 1, user must select Seal Location option code M in Table 6 of Rosemount DP Level PDS. (5) Not recommended for use with spiral wound metallic gaskets (see 1199 product data sheet, document 00813-0100-4016 for additional options).
Page 146
(24) Not available with Housing code 00, 5A, 5J, or 7J. Available with Intrinsically Safe approvals only. For FM Intrinsically Safe, Division 2 (option code I5) or FM FISCO Intrinsically Safe (option code IE), install in accordance with Rosemount drawing 03151-1009 to maintain outdoor rating (NEMA 4X and IP66).
Emerson Process Management GmbH & Co. — Wessling, Germany Emerson Process Management Asia Pacific Private Limited — Singapore Beijing Rosemount Far East Instrument Co., LTD — Beijing, China Emerson Process Management LTDA — Sorocaba, Brazil Emerson Process Management (India) Pvt. Ltd. — Daman, India B.1.2...
Page 148
May 2013 European directive information The EC declaration of conformity for all applicable European directives for this product can be found at www.rosemount.com. A hard copy may be obtained by contacting an Emerson Process Management representative. ATEX directive (94/9/EC) Emerson Process Management complies with the ATEX Directive.
Non-Incendive for Class I, Division 2, Groups A, B, C, and D. Dust Ignition-proof for Class II/III, Division 1, Groups E, F, and G. Ambient temperature limits: -50 to 85° C For use with Rosemount SmartPower options 00753-9220-0001 only. Enclosure Type 4X / IP66 CSA - Canadian Standards Association Process sealing All CSA hazardous approved transmitters are certified per ANSI/ISA 12.27.01-2003.
Page 150
IECEx certifications IECEx Intrinsic Safety Certificate No.: IECEx BAS 04.0017X Ex ia IIC T4 (Ta = -60 °C to 70 °C) For use with Rosemount SmartPower options 00753-9220-0001 only. IP66 Special conditions for safe use (x) The surface resistivity of the antenna is greater than one gigaohm. To avoid electrostatic charge build-up, it must not be rubbed or cleaned with solvents or a dry cloth.
Appendix C: High Gain Remote Antenna Option Reference Manual May 2013 00809-0200-4802, Rev BA Appendix C High Gain Remote Antenna Option Safety messages ............page 143 Functional specifications .
Reference Manual Appendix C: High Gain Remote Antenna Option 00809-0200-4802, Rev BA May 2013 Functional specifications Output WirelessHART 2.4 GHz DSSS (Direct Sequence Spread Spectrum). Radio Frequency Power Output from Antenna: High Gain Remote (WN option) antenna: Maximum of 40 mW (16 dBm) EIRP ...
Reference Manual Appendix C: High Gain Remote Antenna Option 00809-0200-4802, Rev BA May 2013 Vibration 3g Max vibration Installation considerations Antenna mounting Mount antenna vertically (±5°) Antenna height Mount antenna 14 feet (4.3 meters) above infrastructure with clear line of sight. Coaxial Cable Ensure that coaxial cable is securely affixed to the mast to avoid excessive cable movement.
Reference Manual Appendix C: High Gain Remote Antenna Option 00809-0200-4802, Rev BA May 2013 Dimensional drawings Device Connection & RF Lightning Arrestor High Gain Remote Antenna Option...
Reference Manual Appendix C: High Gain Remote Antenna Option 00809-0200-4802, Rev BA May 2013 Installation steps Mount the transmitter following best practice mounting procedures as outlined in the Quick Installation Guide and Product Manual. Connect the RF Lightning protector to the device and tighten. RF Lightning Protector Connect antenna to mounting bracket and tighten nut carefully.
Page 156
Reference Manual Appendix C: High Gain Remote Antenna Option 00809-0200-4802, Rev BA May 2013 Unwind coaxial cable and connect the cable to both the antenna and the lightning protector connected to the transmitter. Leaving one loop minimum for a drip loop. Ensure the drip loop is lower than the device allowing water to flow away from the device.
Page 157
Reference Manual Appendix C: High Gain Remote Antenna Option 00809-0200-4802, Rev BA May 2013 Attach u-bolts to mounting bracket in correct orientation ensuring that antenna will be positioned in a vertical position. Tighten u-bolts to mast and ensure that antenna is pointed in a vertical direction. High Gain Remote Antenna Option...
Page 158
Reference Manual Appendix C: High Gain Remote Antenna Option 00809-0200-4802, Rev BA May 2013 High Gain Remote Antenna Option...
Field Communicator menu trees ......... . page 151 Field Communicator menu trees Figure D-1. Rosemount 3051S Wireless Field Communicator menu tree: Overview Active Alerts 1 Good - only if no alerts 2.
Page 160
Reference Manual Appendix C: Field Communicator Menu Trees 00809-0200-4802, Rev BA May 2013 Figure D-2. Rosemount 3051S Wireless Field Communicator menu tree: Configure Guided Setup 1 Basic Setup 2 Zero Sensor Trim 3 Join Device to Network 4 Configure Update Rate...
Page 161
Reference Manual Appendix C: Field Communicator Menu Trees 00809-0200-4802, Rev BA May 2013 Figure D-3. Rosemount 3051S Wireless Field Communicator menu tree: Manual Setup Broadcast Info Set Range Points 1 Configure Update Rate 1 PV Value 2 Message 1 2 PV Upper Range Value...
Page 162
Reference Manual Appendix C: Field Communicator Menu Trees 00809-0200-4802, Rev BA May 2013 Figure D-4. Rosemount 3051S Wireless Field Communicator menu tree: Manual Setup Active Alert Name Active Alerts 1 Alert Description 1 Refresh Alerts 2 Recommended Actions 2 Active Alert Name #1...
Page 163
Reference Manual Index 00809-0200-4802, Rev BA May 2013 Index ......54 ....30 Analog output trim Environmental considerations .
Page 164
Reference Manual Index 00809-0200-4802, Rev BA May 2013 Mounting Terminal block diagrams ......40 Bolt installation HARTprotocol .
Page 166
Standard Terms and Conditions of Sale can be found at www.rosemount.com/terms_of_sale The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount. the Rosemount logotype, and SMART FAMILY are registered trademarks of Rosemount Inc. Coplanar is a trademark of Rosemount Inc.
Need help?
Do you have a question about the 3051S and is the answer not in the manual?
Questions and answers