Applications; Remote Status And Control Using The Sim Interlock I/O Connector; Figure 6.1 - Remote Status And Control Pinouts; Figure 6.2 - Remote Run/Standby Monitor - AE Techron 7224 Operator's Manual

Single-channel industrial amplifiers for demanding, high-power systems
Hide thumbs Also See for 7224:
Table of Contents

Advertisement

6 Applications

6.1 Remote Status and Control using
the SIM Interlock I/O Connector
The procedures outlined in this section assume
competence on the part of the reader in terms of
amplifier systems, electronic components, and 
good electronic safety and working practices.
Figure 6.1 – Remote Status and Control Pinouts
Figure 6.2 – Remote Run/Standby Monitor
22
7224 & 7226 OPERATOR'S MANUAL – SECTION 6
AE Techron 7224/7226 amplifiers come with a 
SIM-BNC input module that also contains a fe-
male, 25-pin D-Sub connector. This connector can 
be used to provide remote control and monitoring 
of the amplifier. 
The information provided here will instruct you in
the wiring of several control and status applica-
tions including:
Remote Enable / Standby control
Run/Standby status
Over-temperature status
Overload status
Overvoltage status
Reset after Overload error
Voltage monitor
Current monitor
Figure 6.1 maps the pins used for these applications.
For a detailed chart of all DB-25 pinouts, 
see "Appendix 1."
6.1.1 Remote Run/Standby Status
Monitor
Using the SIM-BNC Interlock connector located on 
the back panel of the amplifier, you can remotely 
monitor the Run/Standby status of the amplifier. 
Remote Run/Standby Status
Purpose: Use a voltage meter to monitor the sta-
tus of the amplifier to determine if the amplifier is in 
a "Run" or "Standby" state.
Method: Connect a voltage meter to monitor the 
circuit voltage. Connect across PIN 4 (Interlock) 
and PIN 10 (Sampled Common). 
When the voltage meter reads greater than 10V, 
the amplifier is in the Run state; when the meter 
reads less than 10V, the amplifier is in the Standby 
state. See Figure 6.2.
Signal Type: DC
Level when Asserted: >10 V
Level when Deasserted: <10 V
IMPORTANT: This circuit has a 100K pull-up resis-
tor. Make sure the monitor function has sufficient 
impedance to avoid accidentally influencing status. 
Information subject to change
97-8004167 07-14-17

Advertisement

Table of Contents
loading

This manual is also suitable for:

7226

Table of Contents