X-Ray Tube - Siemens MOBILETT XP Instructions For Use Manual

Mobile x-ray
Hide thumbs Also See for MOBILETT XP:
Table of Contents

Advertisement

Technical Data
Stabilization:

X-ray tube:

Anode rotation:
Anode heat storage
capacity:
Tube housing heat storage
capacity:
Inherent filtration
- Collimator:
- Tube housing assembly:
- Totally:
Collimator:
Collimator light and light
field intensity:
Tube housing assembly:
Leakage technique
1
factors
1
The leakage radiation of the tube housing assembly and attached collimator complies to the requirement of 21 CFR part 1020.3 (K) and
IEC60601-1-3. For further informations see data sheet X-ray tube.
80
of 106
X-Ray Tube
Microprocessor-based adjustment and regulation during exposure
Siemens P135/30 R with nominal focal spot value 0.8 (acc. to IEC 60336)
8 500 r.p.m., 142 Hz (anode angle 15°)
90 000 J = 122 000 HU
800 000 J = 1 100 000 HU
min. 1.6 mm Al (at 70 kVp)
min. 1.5 mm Al (at 70 kVp)
min. 3.1 mm Al according to IEC 60 522
Manually-adjustable multi-leaf collimator, rotatable ±90°
Halogen lamp min. 180 lux at 1 m SID
Maximum temperature 60°C, maximum 2000 mAs/h, nominal continuous
rating 75 W until the covers reach 60°C
133 kVp/10 mAs, 200 exposures/h
Operator Manual
SPR8-230.621.01.03.02