Technical Data
Technology:
Active area:
Operating temperature:
Weight:
Typical filtration:
Warm-up time:
Chamber light transmis-
sion:
Energy dependence:
Reproducibility:
DAP resolution:
Maximal measurable DAP: 1 x 10
Dose rate linearity:
Permissible ambient
temperature:
Permissible relative
humidity:
Permissible air pressure
area:
1
maximum 15 weeks
82
of 106

DAP-equipment (Option)

Ionization chamber, Display and control panel
146 x 146 mm
15 - 40°C
0.24 kg
< 0.4 mm Al equivalent
Ready for use 10 seconds after power on. Specified accuracy achieved 15
minutes after power on.
≥ 70 %
50 - 150 kV ±8 %, related to 100 kV (IEC 60 580)
40 - 50 kV ±3 %, related to 50 kV
> 99 % under constant pressure and temperature
2
0.1 µGym
2
(0.01 µGym
when using a system with high sensitivity)
6
2
µGym
5
2
(1 x 10
µGym
when using a system with high sensitivity)
< 2 % deviation

Environmental conditions

Operation
+10°C ... +40°C
30 % ... 75 %
700 hPa ... 1060 hPa
Transport/Storage
-20°C ... +70°C
1
10 % ... 100 %
SPR8-230.621.01.03.02