Siemens MOBILETT XP Instructions For Use Manual page 70

Mobile x-ray
Hide thumbs Also See for MOBILETT XP:
Table of Contents

Advertisement

The Exposure Point System
Example 1
Example 2
68
of 106
Doubling the density using mAs
The density in the following example will be doubled by pressing the mAs+
button six times.
70 kV
20 mAs
Doubling the density using kV
Adjusting the kV will affect both density and contrast. The density in the following
example will be doubled by pressing the kV+ button three times (whole exposure
points).
70 kV
20 mAs
70 kV
40 mAs
81 kV
20 mAs
Operator Manual
SPR8-230.621.01.03.02