Exposure Points
MOBILETT
SPR8-230.621.01.03.02
Setting the Exposure Parameters
The exposure parameters are set by using the kV and mAs buttons.
For more detailed information see chapter Technical Data.
The exposure data are listed in chapter Dosimetry
(→ Page 93)
The concept behind the exposure point system is to easily change the density or
contrast of the film. Therefore, the kV and mAs may appear to change in odd
steps. This will, however, allow logical density and contrast adjustments.
Exposure points are set by Siemens customer service. Examples can be found in
chapter Dosimetry
(→ Page 93)
Selecting exposure parameters
◆ Press the kV or mAs buttons briefly.
– The exposure value increases/decreases by one step.
◆ Hold the button pressed down for a period of time.
– This results in a continuous increase/decrease of the particular exposure
parameter.
Operation
43
of 106