Siemens MOBILETT XP Instructions For Use Manual page 80

Mobile x-ray
Hide thumbs Also See for MOBILETT XP:
Table of Contents

Advertisement

Technical Data
mAS-range (Tube current
and exposure time
product)
Mains operation 110 V
MOBILETT XP / XP Hybrid
mAS-range (Tube current
and exposure time
product)
Mains operation 110 V -
230 V
MOBILETT XP variant
ECO
mAs-range: (Tube current
and exposure time
product)
Battery Operation
MOBILETT XP Hybrid only
mAs accuracy:
Shortest exposure time:
Max. exposure time:
Overload protection:
Length of power cable:
Exposure release:
Weight:
78
of 106
At 40 - 55 kV:
At 57 - 60 kV:
At 63 - 70 kV:
At 73 - 77 kV:
At 81 - 85 kV:
At 90 - 96 kV:
At 102 kV:
At 105 - 117 kV:
At 125 - 133 kV:
At 40 -52 kV:
At 55 -60 kV:
At 63 -66 kV:
At 70 -73 kV:
At 77 -81 kV:
At 85 -90 kV:
At 96 -102 kV:
At 109 -117 kV:
At 125 kV:
At 40 - 81 kV:
At 83 - 90 kV:
At 96 - 102 kV:
At 109 - 117 kV:
Up to 133 kV:
0.32 mAs - 2 mAs ± 0.1 mAs
> 2mAs ± 5 %
1 ms for MOBILETT XP / XP Hybrid
2 ms for MOBILETT XP ECO
5.0 s for MOBILETT XP Hybrid
3.5 s for MOBILETT XP
3.0 s for MOBILETT XP ECO
Heat units continuously calculated and monitored by the microprocessor
6 m
release switch with 5 m (16.4 ft) coiled cord, optional IR remote control
Approx. 285 kg, MOBILETT XP Hybrid
Approx. 210 kg, MOBILETT XP
mAs 0.32 - 140
mAs 0.32 - 125
mAs 0.32 - 110
mAs 0.32 - 100
mAs 0.32 - 90
mAs 0.32 - 80
mAs 0.32 - 71
mAs 0.32 - 63
mAs 0.32 - 56
mAs 0.5 -125
mAs 0.5 -110
mAs 0.5 -100
mAs 0.5 -90
mAs 0.5 -80
mAs 0.5 -71
mAs 0.5 -63
mAs 0.5 -56
mAs 0.5 -50
mAs 0.32 - 360
mAs 0.32 - 320
mAs 0.32 - 280
mAs 0.32 - 250
mAs 0.32 - 220
SPR8-230.621.01.03.02