High Priority Alarms - Inogen at home User Manual

Hide thumbs Also See for at home:
Table of Contents

Advertisement

High Priority Alarms

The following high priority alarm messages are accompanied by a five beep
pattern, repeated every 10 seconds and a flashing yellow light warning
indicating immediate response by the operator is required.
Indicator
5
4
3
2
p,
Sound indicator
only
NOTE When power is lost during operation, the buzzer will alarm
Condition
Fire Hazard
System Hot
System Cold
System Error
Power Error
continuously.
Action/ Explanation
Smoke has been detected inside the device and
device is shutting down. Allow the concentrator
to air out in a smoke-free environment and
restart. If condition persists, switch to a backup
source of oxygen and contact your equipment
provider.
Concentrator temperature is too high for
operation. Ensure air intake and outlet vents
have clear access and particle filters are clean.
Allow the concentrator to cool for 10 minutes
and restart. If condition persists, switch to a
backup source of oxygen and contact your
equipment provider.
Concentrator temperature is too cold for
operation. Allow the concentrator to warm
up in a room temperature environment for 10
minutes and restart. If condition persists, switch
to a backup source of oxygen and contact your
equipment provider.
Switch to a backup source of oxygen and
contact your equipment provider.
Concentrator lost power during operation.
Try unplugging the power cord and plugging
it back in. If alarm does not reset, press and
hold the power button down for 2 seconds. If
condition persists, switch to a backup source of
oxygen and contact your equipment provider.
9

Advertisement

Table of Contents
loading

Table of Contents