Siemens SITRANS F MASS 6000 Ex d Operating Instructions Manual page 6

Coriolis flowmeters
Hide thumbs Also See for SITRANS F MASS 6000 Ex d:

Advertisement

SITRANS F C MASSFLO
Horizontal mounting
Vertical mounting
Vibration
Cross talk
Zero-point adjustment
A5E02944883-A
Transmitter MASS 6000 Ex-d may
Mount the sensor on a vibration-free wall or
steel frame as shown.
Ensure that the sensor is not emptied of liquid
(during normal operation) otherwise incorrect
measurement will occur.
Locate the sensor low in the system in order to
avoid an under-pressure in the sensor sepera-
ting air/gas in the liquid.
possible from components that generate
mechanical vibration in the piping.
Cross talk between sensors mounted close to
each other may disturb the measurement. To
avoid cross talk never mount more than one
connections between the sensors as shown.
with tight shut-off should always be mount-
ed in connection with the sensor as a proper
accuarcy.
Please read and follow the directions stated in
the instruction SFIDK.PI.028.P1.52 regarding
build-in of sensor.
The instruction is supplied with the sensor.