Download Print this page

Approvals - Siemens SITRANS F MASS 6000 Ex d Operating Instructions Manual

Coriolis flowmeters
Hide thumbs Also See for SITRANS F MASS 6000 Ex d:

Advertisement

SITRANS F C MASSFLO
Material
Rating
Load
upply voltage
Power consumption
Power supply
x approvals
Intrinsically safe sensor
interface
Cable between sensor and
MASS 6000
(continued)
Stainless steel AISI 316 W1.4435
Compact mounted on sensor, IP 67 (NEMA 4X) to IEC 60529
Remote mounted, IP 65 (NEMA 4) to IEC 60529
Emission
EN 55011-1 (Light industry)
Immunity
EN EN 61326-1 (Industry)
Within the value limits according to "Allgemeine Anforderung" with error criteria A in accordance with NE 21
24 V AC/DC
6 VA I
= 250 mA, I
= 2A (30 msec.)
N
ST
The power supply shall be from a safety isolating
transformer. Cable core is 0.2 to 2.5 mm
Ex d e ib [ia Ga] IICT4 Gb, DEMKO 03 ATEX 135253X
[V]
o
I
[mA]
o
P
[W]
o
L
or
[mH]
o
L
/R
[ H/ ]
o
o
C
[nF]
o
Maximum distance between sensor and transmitter MASS 6000 is 300 m with Siemens Flow
Instruments cable FDK:083H3005 or FDK:083H3006 or equivalent cable.
Cable data
Capacitance
Self-inductance
Resistance
Maximum length
C
max.
L
/R
c
c
Isolation
Temperature range
A5E02944883-A
24 V AC/DC
6 VA I
The power supply shall be from a safety isolating
transformer. Cable core is 0.2 to 2.5 mm
2
16
85
0.51
0.022
5 or
70
150
1410
= 250 mA, I
= 2A (30 msec.)
N
ST
12
7.3
0.018
682
1410
300 [pF/m]
1
H/m]
0.05 [ /m]
300 [m] / 984 [ft]
100 [nF]
25 [ H/ ]
0.25 [mm]
20 C to +105 C/ 4 F to +221 F
2
12
6
998
17

Advertisement

loading