Pneumatic Diagram; Unit Specifications; Provider's Notes; Viii. Pneumatic Diagram - DeVilbiss PD1000 Service Manual

Compact conversing device
Hide thumbs Also See for PD1000:
Table of Contents

Advertisement

P N E u M AT I C D I A G R A M / u N I T s P E C I F I C AT I O N s

VIII. PNEuMATIC DIAGRAM

IX. uNIT sPECIFICATIONs
Weight
Dimensions
Power Supply
Operational Voltage Range
Operating Temperature Range
Operating Pressure Range
Operating Atmospheric Conditions
Operating Humidity Range
Storage and Transportation Temperature Range
Storage and Transportation Humidity Range
Degree of Protection Against Ingress of liquids
Degree of Protection Against Electric Shock
Power Requirements
Expected Shelf and Service Life (excluding batteries)
Modes of Operation
Approval Body And Safety Standards
US Patents
X. PROVIDER's NOTEs
No routine calibration or service is required provided the device is used in accordance with the manufacturer's
directions. Between patients wipe with a damp cloth having a maximum 5.25% Sodium Hypochlorite (Bleach) or 3%
Hydrogen peroxide solution. Avoid getting fluids or debris such as sand or dirt inside the oxygen connections. Do not
immerse in water.
14. 7 ounces (16.3 ounces with battery)
4.75"L x 3.4"W x 2.8"H (12.06 cm L x 8.64 cm W x 7.11 cm H)
(2) Standard "AA" alkaline or NiMH. NOTE-Batteries other than alkaline
or NiMH are not recommended due to their limited capacities.
2.3 to 3.6V DC
5° to 40° C (41˚ to 104°F)
500 to 2250 PSIG (34 to 155 Bar) tank pressure
500 to 1020 millibar
0 to 95% R.H., non-condensing
20° to 60° C ( -4° to 140° F)
Up to 95% R.H., Non-condensing
None
TYPE BF applied part
Average steady state "ON" current 1.6 uA. Batteries other than alkaline
or NiMH are not recommended due to the capacity needed for operation
and battery life of the unit. Typical new battery life is 200 hours when
used at 25°C, 2 LPM and 20 BPM. Settings and breath rate will affect
battery life. After the Low Battery (flashing red) light illuminates, the unit
will continue to operate about four hours when used at 25°C, 20 BPM
and the 6 LPM setting. Settings, breath rate, and battery conditions will
affect use times. Refer to local regulations for battery recycling and/or
disposal requirements
5 years based on 4 hours use per day at 20 BPM
Continuous/Pulsed
IEC 601-1;CAN/CSA-C22.2 No. 601.1-M90 and IEC 601-1-2
4,519,387; 5,755,224; 4,457,303
9
LT-1859

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents