Important Control Solution Information
Use only FORA control solutions.
Check the expiration date on the control solution vial. Do not use if expired.
It is recommended to use the control solutions between the temperatures of
20℃ and 25℃ (68℉ and 77℉.)
Shake the vial, discard the first drop of control solution, and wipe off the
dispenser tip to ensure a good sample and an accurate result.
Use only within 3 months after first opening. Record the discard date (date
opened plus 90 days) on the control solution vial. Discard on this date.
Store the control solution tightly closed at temperatures 2℃ - 30℃ (36℉ -
86℉). Do not freeze.
NOTE
The control solution range printed on the test strip vial is for FORA control
solution only. It is used to test monitor and test strip performance. It is not the
recommended range for your blood glucose level.