Important Control Solution Information - ForaCare GD40 Owner's Manual

Table of Contents

Advertisement

Available languages

Available languages

Important Control Solution Information

• Use only ForaCare control solutions with your meter.
• Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution
vial and discard the remaining solution after 3 months.
• It is recommended that the control solution test be done at room
temperature 68 ° F to 77 ° F (20 ° C to 25 ° C). Make sure your control
solution, meter, and test strips are at this specified temperature
range before testing.
• Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an ac-
curate result.
• Store the control solution tightly closed at temperatures between
35.6 ° F to 86 ° F (2 ° C to 30 ° C). Do NOT freeze.