Important Control Solution Information

● Use only our control solutions with your meter.
● Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution bottle
and discard the remaining solution after 3 months of first opening.
● It is recommended that the control solution test be done at room
temperature (68°F to 77°F / 20°C to 25°C). Make sure your control
solution, meter, and test strips are at this specified temperature range
before testing.
● Shake the bottle before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
● Store the control solution tightly closed at temperatures between
35.6°F and 86°F (2°C and 30°C). Do NOT freeze.
32