USER'S MANUAL FA-405
Explain:
When the message of A<cr>is sent to the power supply from computer,
the power supply will return the following message immediately:
The contents consist of the uppercase A, the numeral from 0 to 9 and
decimal. Further details is described as follows:
W
Function: The present output load, the unit is W.
W<cr>
Syntax:
Explain:
When the message of W<cr>is sent to the power supply from computer,
the power supply will return the following message immediately:
The contents consist of the uppercase W, the numeral from 0 to 9 and
decimal. Further details is described as follows:
U
Function: The maximum voltage limit at present, the unit is V.
U<cr>
Syntax:
Explain:
When the message of U<cr> is sent to the power supply from computer,
the power supply will return the following message immediately:
The contents consist of the uppercase U and the numeral from 0 to 9.
Further details is described as follows:
When the uppercase U becoming the lowercase u means that the power
supply is in the setting status of voltage limit mode.
12/2003
Aa.aaa<cr> 6 characters totally + CR/LF
a.aaa = The present output current, the unit: A
HEX = 57 0D
Wwww.w<cr> 6 characters totally + CR/LF
www.w = The present output load, the unit: W
HEX = 55 0D
Uuu<cr> 3 characters totally + CR/LF
uu = The maximum voltage limit at present, the unit: V
Page 21