2.2 Installation of SONOST 3000

NOTE !
Since SONOST3000 is made up of precise components, you should install the
product according to instructions below.
Do not install and keep SONOST3000 at
excessively high or low temperatures.
Proper
temperature:
18~27 ℃
Do not install the system in the place with
polluted air and high humidity and do not
expose the machine to direct ray sunlight. It
is advisable to use SONOST3000 in the
place equipped with air conditioning or
heating.
Proper humidity:
20~80%
Do not share the power outlet with other
products through an extension cord.
Do not cover ventilation of the system or
place it near the wall. High inside
temperature of the machine might cause a
fire.
OT00-2F0423-01
DOC. No. : OT06-2R0422-00-MUL
Do not install or keep SONOST3000 in
the place where the machine can be
rocked or shaken. Make sure to place
the system
parallel to the
floor.
Install SONOST3000 in the place where
it is free from water or chemicals.
For escaping from electronic noise, do not
install near a electronic generator, X-ray
equipment, broadcasting equipment.
It causes a
result to be
inaccurate.
100-240V is usable. When SONOST3000
is taken out of a warehouse, insulating
transformers should be set to 100-240V.
Since sudden power outage can remove
all data saved in the
product, you should
install
SONOST3000 in the
place where a power
supply is stabilized.