3.2.7 Calibration of SONOST 3000(DailyTest)

 DailyTest
Make sure to calibrate the system to reduce a margin of error and to get
accurate value in times of examining bone density. It is to raise credibility of
examination results through checking of system reliability on a daily basis.
To calibrate the system, click the <DailyTest>button in the "Home" screen.
Then the "DailyTest" screen appears as shown below. When you click the
<StartDailyTest>button, DailyTest is started. If the DailyTest success, the
message box "Complete Dailytest" appears and back to "Home" screen.
If you select the "Time" tap, graph appears in the time domain.
If you select the "Frequency" tap, graph appears in the frequency domain.
Whenever you calibrate the system, you should use the QC Phantom
included in the accessories list.
If you click the <Start>button on the bottom, ultrasonic waves are generated
and the word on the <Start> button is changed into <Stop>. In case of
emergency, click this button and you can stop daily testing. In the normal
condition, the message box "Complete Dailytest" appears as shown below with
an alarm sound after progressing calibration of the system and if you click
<OK>button, the progress of calibration is completed.
OT00-2F0423-01
DOC. No. : OT06-2R0422-00-MUL
DailyTest Screen