Introduction Congratulations on your purchase of the Extech EX505 True RMS Autoranging Multimeter. This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency, Diode Test, and Continuity plus Thermocouple Temperature. It features a waterproof, rugged design for heavy duty use This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service.
Page 3
• Expired or damaged batteries can cause cauterization on contact with the skin. Always, therefore, use suitable hand gloves in such cases • See that the batteries are not short-circuited. Do not throw batteries into the fire. EX505-EU-EN-V7.0-12/11...
NEVER operate the meter unless the back cover and the battery and fuse covers are in place and fastened securely. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. EX505-EU-EN-V7.0-12/11...
(10 ) (volts, amps) Amps kilo (10 ) (ohms) Farads (capacitance) mega (10 ) (ohms) Ω Ohms Hertz (frequency) Volts Percent (duty ratio) Relative Alternating current Auto Autoranging Direct current HOLD Display hold ºF Degrees Fahrenheit ºC Degrees Centigrade EX505-EU-EN-V7.0-12/11...
Touch the red test probe tip to the “hot” side of the circuit. Read the voltage in the display. Press the HZ/% button to indicate “Hz”. Read the frequency in the display. Press the Hz/% button again to indicate “%”. Read the % of duty cycle in the display. EX505-EU-EN-V7.0-12/11...
10. Press the Hz/% button to indicate “Hz”. 11. Read the frequency in the display. 12. Press the Hz/% button again to indicate “%”. 13. Read the % duty cycle in the display. 14. Press the Hz/% button to return to current measurement. EX505-EU-EN-V7.0-12/11...
V on the display. Touch the test probes to the diode under test. Forward voltage will typically indicate 0.400 to 0.700V. Reverse voltage will indicate “OL”. Shorted devices will indicate near 0V and an open device will indicate “OL” in both polarities. EX505-EU-EN-V7.0-12/11...
Hz jack. Touch the test probe tips to the circuit under test. Read the frequency on the display. Press the Hz/% button again to indicate “%” on the display. Read the % of duty cycle on the display. EX505-EU-EN-V7.0-12/11...
The auto off feature will turn the meter off after 30 minutes. To disable the auto power off feature, hold down the MODE button and turn the meter on. LOW BATTERY INDICATION icon will appear in the display when the battery voltage becomes low. Replace the battery when this appears. EX505-EU-EN-V7.0-12/11...
USE ONLY FRESH BATTERIES OF THE RECOMMENDED SIZE AND TYPE. Remove old or weak batteries so they do not leak and damage the unit. IF THE METER IS TO BE STORED FOR A LONG PERIOD OF TIME, the batteries should be removed to prevent damage to the unit. EX505-EU-EN-V7.0-12/11...
[SIBA 70-172-40], 10A/1000V fast blow for the 20A range [SIBA 50-199-06]). Replace and secure the rear cover, battery and battery cover. WARNING: To avoid electric shock, do not operate your meter until the fuse cover is in place and fastened securely. EX505-EU-EN-V7.0-12/11...
400μA 0.1μA 4000μA 1μA ±(2.5% reading + 5 digits) 40mA 0.01mA 400mA 0.1mA 0.01A ±(3.0% reading + 7 digits) NOTE: Accuracy is stated at 18 C to 28 C (65 F to 83 F) and less than 75% RH. EX505-EU-EN-V7.0-12/11...
Page 14
(probe accuracy not included) -20 to 750°C Note: Accuracy specifications consist of two elements: • (% reading) – This is the accuracy of the measurement circuit. • (+ digits) – This is the accuracy of the analog to digital converter. EX505-EU-EN-V7.0-12/11...