10 Diagnostics and servicing
Failure
Code
Reason
Text message
F013
Sensor does not detect an echo
during operation
no measured val-
Process component or probe
ue available
contaminated or defective
F017
Adjustment not within specification
Adjustment span
too small
F025
Index markers are not continu-
ously rising, for example illogical
Error in the line-
value pairs
arization table
F036
Failed or interrupted software
update
No operable soft-
ware
F040
Hardware defect
Error in the elec-
tronics
F041
Cable probe broken or rod probe
defective
Probe loss
F080
General software error
General software
error
60
This status message is always active. It cannot be deactivated by the
user.
Function check: The instrument is being worked on, the measured
value is temporarily invalid (for example during simulation).
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
Out of specification: The measured value is unreliable because an
instrument specification was exceeded (e.g. electronics temperature).
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
Maintenance: Due to external influences, the instrument function
is limited. The measurement is affected, but the measured value is
still valid. Plan in maintenance for the instrument because a failure is
expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
The following table shows the error codes in the status message
"Failure" and gives information on the reason and rectification. Keep
in mind that some information is only valid for four-wire instruments.
Rectification
Check for correct mounting and/or
parameter settings
Clean or exchange process com-
ponent or probe
Change adjustment according
to the limit values (difference
between min. and max. ≥ 10 mm)
Check values of the linearization
table
Delete/create a new linearization
table
Repeat software update
Check electronics version
Exchanging the electronics
Send instrument for repair
Exchanging the electronics
Send instrument for repair
Check probe and exchange, if
necessary
Disconnect operating voltage
briefly
VEGAFLEX 82 • 4 ... 20 mA/HART - two-wire
DevSpec State
in CMD 48
Bit 0 of
Byte 0 ... 5
Bit 1 of
Byte 0 ... 5
Bit 2 of
Byte 0 ... 5
Bit 3 of
Byte 0 ... 5
Bit 4 of
Byte 0 ... 5
Bit 13 of
Byte 0 ... 5
Bit 5 of
Byte 0 ... 5