HCU2 error reporting

The HCU2 LCD display gives information on head and system errors.

Datum error

Fault
The motorised probe head has not locked up to the required level of accuracy.
Action
Refer to your motorised head user's guide for further information concerning this error condition.

Obstruct error

Fault
The motorised probe head has been obstructed while rotating to the required position.
Action
Refer to your motorised head user's guide for further information concerning this error condition.

Overload error

Fault
The motorised probe head head has been overloaded during a machine quill move.
Action
Refer to your motorised head user's guide for further information concerning this error.
Head not present
Fault
The motorised head is not electrically connected to the probe head controller.
Action
Check the head is electrically connected.
Check cable integrity.
Contact your CMM supplier for further assistance.
HCU2 not recognised
Fault
There is a communications error between the HCU2 and the motorised head controller.
Action
Check cable connections.
Contact your Renishaw representative for further assistance.