Maintenance 
 Keep the meter's measurement electrode clean. Between measurements, rinse the electrode 
with deionized or distilled water. If the electrode has been exposed to a solvent immiscible in 
water, clean it with a solvent miscible in water, e.g. Ethanol, and then rinse carefully with 
water. 
 Store the electrode carefully. Before storing, rinse carefully in deionized or distilled water and 
store dry. 
Troubleshooting 
Power ON is attempted but there is no display 
 Be sure to press the ON‐OFF button for at least 100mS to switch the meter ON. 
 Check that batteries are positioned correctly, making good contact, and correct polarity. 
 Replace the batteries if necessary. 
 Remove and replace the existing batteries. 
Display switches OFF 
 This is normal when Auto Power OFF is activated. 
 Replace the batteries if necessary. 
Air Bubbles adhered to Electrode 
 Stir the electrode completely and be sure to dip the electrode into a solution at an oblique 
angle. Vertical dipping can cause many air bubbles to adhere. 
 Gently tap the bottom of the solution container while stirring the electrode in the solution. 
 Air can be blown across the electrode before dipping it into the solution. 
Error Codes 
 Refer to the Table below for details on Error Codes displayed by the meter.  
Code  Description 
SALINITY ERRORS 
‐‐‐‐‐ 
E03 
E04 
TEMPERATURE ERRORS 
E01 
E02 
E03 
nd Quality Products Online at:
Measurement outside of 
range 
Salinity is over‐range 
Temperature error 
Temperature circuit 
damaged 
Temperature value is below 
allowable range or 
Temperature circuit damage 
Temperature value is above 
allowable range or 
Temperature circuit damage 
GlobalTestSupply
www.
Suggestions 
In Manual Range mode, press and hold the up 
arrow for 2 seconds to change range or use the 
Auto Range mode 
Check against a standard buffer solution. If 
problem persists, repair meter 
Repair meter 
Repair meter 
Check again at room temperature. If error 
persists, repair meter 
Check again at room temperature. If error 
persists, repair meter 
9
.com