How WP4C takes Readings

WP4C's cooled mirror is controlled at the chamber dew
point and its temperature is measured.. When the instru-
ment has finished its read cycle, the water potential is dis-
played, accompanied by the LED flash and beeper (if you
have the beeper enabled).

Cautions

Never leave a sample in your WP4C after a reading has
finished. The sample may spill and contaminate the
instrument's chamber if the instrument is accidentally
moved or jolted.
Never try to move your instrument after a sample has
been loaded. Movement may cause the sample mate-
rial to spill and contaminate the sample chamber.
Take special care not to move the sample drawer too
quickly when loading or unloading liquid samples, in
order to avoid spilling.
If a sample has a temperature that is higher than the
WP4C's chamber (Ts-Tb is a positive number), take
the sample out immediately, put a cap on it, and cool
it. Warm samples can cause condensation in the
chamber and adversely affect subsequent readings.