Setting The Temperature - Decagon Devices WP4C Operator's Manual

Dewpoint potentiameter
Table of Contents

Advertisement

WP
PotentiaMeter
4C
4. The Menus

Setting the Temperature

The WP4C gives you the ability to manually set your
instrument's sample chamber temperature. This is done by
set T
pressing the lower right button next to the "
" in
the system configuration menu. The following screen will
appear:
+
Adjust setpoint
-
-Exit-
25.0
Adjusting the Setpoint Temperature
Use the buttons next to + and - to adjust the target set-
point temperature (displayed in the lower right corner). If
you press either button it adjusts in increments of .1°C.
Note: Holding down the button will rapidly increment the value.
The target setpoint temperature roughly corresponds to
the temperature at which you wish the sample to read.
Therefore, adjust the setpoint to the temperature that you
want, then begin measurements to see how close your
WP4C comes to your desired temperature (this works best
by putting the WP4C in continuous mode). After several
samples, it should show consistent temperature readings.
At this point, make any needed adjustment to the setpoint
index number to reach your desired temperature. You will
be able to adjust the index number between 15° and 50°C.
If you press the - button after you reach 15°, it will disable
18