Dry Samples; Samples And Temperature - Decagon Devices WP4C Operator's Manual

Dewpoint potentiameter
Table of Contents

Advertisement

restrict water transfer. For short-term storage (< 3
hours) the cup lid is acceptable. If it will be a long time
before the measurement is made, seal the lid with tape
or Parafilm completely around the cup/lid junction.

Dry Samples

Samples that have a water potential drier than about -300
MPa cannot be accurately measured with the WP4C.
However, samples with such dry water potential values are
rare. When a sample's water potential value is drier than
about -300 MPa, WP4C will display an error message indi-
cating the lowest reading it could make on that particular
sample. For example, if you were measuring a dry sample
and the following screen appeared:
this screen indicates that the last water potential reading
the WP4C measured on this sample was -301.8 Megapas-
cals. Therefore, the actual water potential of the sample is
lower than the instrument can measure.

Samples and Temperature

If samples are warmer than the chamber when they are
placed in it (Ts-Tb is a positive number), condensation
may occur and moisture may condense inside the block. In
order to prevent this, do the following:
< -820 MPa
27
WP
PotentiaMeter
4C
6. Sample Preparation

Advertisement

Table of Contents
loading

Table of Contents