5. Calibration and Verification
5. Calibration and Verification
Verification
WP4C uses the chilled mirror dewpoint technique for mea-
suring water potential. This is a primary method but cleanli-
ness of the instrument can affect the calibration. The
calibration slope is fixed during factory calibration. The user
just adjusts the zero offset. Calibration can be done with
any solution of known water potential. We recommend
using the 0.5 molal KCl verification standard from Decagon.
Verification Standards
Verification standards are specially prepared salt solutions
that have a specific molality and water potential. The
potassium chloride (KCl) verification standards are accu-
rate, easy to use, and readily available from Decagon
Devices. Most importantly, they greatly reduce prepara-
tion errors. We therefore recommend using Decagon's
KCl Performance Verification Standard for the most accu-
rate verification of your WP4C's performance.
The standards are produced under a strict quality assur-
ance regime and are shelf stable for one year. If for some
reason you cannot obtain Decagon's verification standards
and need to make a salt solution for verification, refer to
Appendix A.