Microlife WatchBP Home A Instructions Manual page 27

Hide thumbs Also See for WatchBP Home A:
Table of Contents

Advertisement

Error
Description
«ERR 1» Signal too
weak
«ERR 2» Error signal
«ERR 3» No pressure in
the cuff
«ERR 5» Abnormal
result
WatchBP Home A
Potential cause and remedy
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.
During the measurement, error
signals were detected by the cuff,
caused for instance by movement or
muscle tension. Repeat the measure-
ment, keeping your arm still.
An adequate pressure cannot be
generated in the cuff. A leak may
have occurred. Replace the batteries
if necessary. Repeat the measure-
ment.
The measuring signals are inaccurate
and no result can therefore be
displayed. Read through the checklist
for performing reliable measurements
and then repeat the measurement.*
Error
Description
«HI»
Pulse or cuff
pressure too
high
«LO»
Pulse too low The pulse is too low (less than 40
* Please contact Microlife, if this or any other problem occurs repeat-
edly.
If you think the results are unusual, please read through
the information in this instruction manual carefully.
Potential cause and remedy
The pressure in the cuff is too high
(over 300 mmHg) OR the pulse is too
high (over 200 beats per minute).
Relax for 5 minutes and repeat the
measurement.*
beats per minute). Repeat the
measurement.*
27