4. Adjust span calibration:
• Scroll to the TWSPAN parameter.
• Press , 000000 will display.
• Set the parameter to match the reading from the multimeter.
• Press
or
to select the digit.
• Press
or
to increment or decrement the value.
TARE
• Press
to move to the decimal point entry.
• Press
or
to adjust the decimal point placement.
TARE
• Press
to accept the displayed value.
5. Verify calibration:
• Return to the TWZERO/TWSPAN parameter and verify that the calibration has not drifted.
• Repaet calibration if needed.
6. Return to normal mode. Analog output function can be verified using test weights.