SCALE Menu
Parameter
Choices
PWRUPM
GO
DELAY
ACCUM
OFF
ON
THRESH
0-999999
CALIBR
WZERO
WVAL
WSPAN
WLIN
REZERO
LAST
TEMP
Power up mode.
GO - In GO mode, the indicator goes into operation immediately after a brief power up
display test.
DELAY - the indicator performs a power up display test, then enters a 30-second warm up
period. If no motion is detected during the warm up period, the indicator becomes
operational when the warm up period ends; if motion is detected, the delay timer is reset
and the warm up period repeated.
Accumulator – specifies if the scale accumulator is enabled or disabled. If enabled, accumulation
occurs every time a print operation is performed, as long as the weight returns to a value lower
than the value set in the THRESH parameter.
Accumulator ResetTheshold – when the weight fall below the value set, the accumulator is reset
Calibration – See Figure 3-10 for descriptions, and see Section 4.0 for calibration procedures.
Table 3-5. Scale Menu Parameters (Continued)
Description
Configuration
29