4. Activate the generator by pressing the appropriate button:
To Activate...
Monopolar
Cut or Blend Modes
Coagulation or Fulguration Modes
Bipolar
Any Bipolar
NOTICES:
One footswitch activates either monopolar or bipolar footswitching accessories.
The Aaron 1250 ™ Electrosurgical Generator powers up to the Cut and Coag modes and their last
set power settings. For example, you select the Cut mode and set the power to 50 watts. When
you turn the unit off, it will default to Cut mode at 50 watts when its turned on again. The
Coagulation mode will also default to the last set power setting. For the Blend, Fulguration or
Biploar modes and power settings, the unit will remember the last set power once the mode is
again selected. For example, you select the Blend mode and set the power to 40 watts and select
the Bipolar mode and set the power to 10 watts. When you turn the unit off, it will default to the
Cut and Coag modes and their last selected power settings when its turned on again. If you then
select Blend, the power will automatically default to 40 watts since it was your last set power
setting.
4-8
Press This...
Yellow Button
Yellow Pedal
Blue Button
Blue Pedal
Blue (Coag) Pedal
On This Device...
Handswitching Pencil
Footswitch
Handswitching Pencil
Footswitch
Footswitch