OUTPUT CHARACTERISTICS
Maximum Output for Monopolar and Bipolar Modes
Power readouts agree with actual power into rated load to within 20% or 5 watts, whichever is greater.
Mode
Output Power
120 W @ 500 Ω
Cut
90 W @ 800 Ω
Blend
80 W @ 1000 Ω
Coagulation
40 W @ 1000 Ω
Fulguration
30 W @ 200 Ω
Bipolar
* an indication of a waveform's ability to coagulate bleeders without a cutting effect
A-8
Output Frequency
357 kHz ± 50 kHz
357 kHz ± 50 kHz
475 kHz ± 19 kHz
410 kHz ± 50 kHz
520 kHz (-14 kHz, +29 kHz)
Open Circuit
Repetition Rate
Vpeak max
N / A
1250V
30 kHz ± 5 kHz
1850V
57 kHz ± 5 kHz
3300V
25 kHz ± 5 kHz
3900V
32 kHz ± 5 kHz
1200V
Bovie Medical Corporation
Crest Factor*
(Rated Load)
2.9 ± 20%
3.3 ± 20%
5.5 ± 20%
7.7 ± 20%
6.9 ± 20%