Setting Hypoglycemic Warning Alarms - Sanofi-aventis BG-STAR Owner's Manual

Hide thumbs Also See for BG-STAR:
Table of Contents

Advertisement

15
Setting
Hypoglycemic Warning Alarms
The hypoglycemic Warning Alarm alerts you when your
glucose test results are below the hypoglycemic threshold
you have set. The meter is pre-set to a hypoglycemic
threshold of 3.9 mmol/L. Always consult your healthcare
professional for your recommended target ranges. When
your glucose test results are below your set hypoglycemic
threshold, the meter display will flash your test results, the
abbreviation "HYPO," time, and date repeatedly.
WARNING:Your healthcare professional should recommend
what your hypoglycemic limit is.
[1] When your meter is deactivated, press the CHECK ( )
button once to enter menu mode. The word "MEM" is
displayed.
[2] Press the DOWN ( ) or UP ( ) arrow buttons
until the abbreviation "HYPO" is displayed.
[3] Press the CHECK ( ) button to enter and edit
the Hypoglycemic Warning Alarm settings.
[4] Set the Hypoglycemic Alarm ON or OFF with
the DOWN ( ) or UP ( ) arrow buttons. Your
selection will be highlighted by the blinking
of the symbol. Press the CHECK ( ) button
to confirm your selection and move on to the
hypoglycemic threshold setting.
44
[5] Set the hypoglycemic threshold
(flashing) with the DOWN ( ) or UP ( )
arrow buttons. Press the CHECK ( ) button
to confirm your selection and return to the
menu.
16
Setting
Hyperglycaemic Warning Alarms
The Hyperglycemic Warning Alarm alerts you when your
glucose test results are above the hyperglycemic threshold
you have set. The meter is pre-set to a hyperglycemic
threshold of 13.3 mmol/L. Always consult your healthcare
professional for your recommended target ranges. When
your glucose test results are above your set hyperglycemic
threshold, the meter display will flash your test results, the
abbreviation "HYPr," time, and date repeatedly.
WARNING: Your healthcare professional should recommend
what your hyperglycemic limit is.
[1] When your meter is deactivated, press the CHECK ( )
button once to enter menu mode. The word "MEM" is
displayed.
[2] Press the DOWN ( ) or UP ( ) arrow buttons
until the abbreviation "HYPER" is displayed.
[3] Press the CHECK ( ) button to enter and edit
the Hyperglycemic Warning Alarm settings.