[4] Viewing Control Solution Test Result:
The result will appear on the display and, if
the volume is enabled, your meter will beep.
The date and the abbreviation "Ctrl" will also
appear on the display.
Compare the result of your control solution
test to the range printed on the BGStar Test
Strip vial label for the control solution level
(Normal or High) that you used. Your control
solution result should fall within this range.
The control solution test result is stored in
memory.
To deactivate the meter, remove the
used BGStar Test Strip. Dispose of
used materials properly.
If your control solution test results are
out of the range, repeat the test. If the
results from the control solution tests
continue to fall outside that range, do
not use the BGStar System to test your
blood glucose. Call Customer Care.
IMPORTANT: Out-of-range test results may be due to one or
more of the following factors:
[1] Expired or defective control solution.
[2] Expired or defective test strip.
[3] Error in performing test.
20
Normal: 4.4-6.7 mmol/L
FOR EXAMPLE ONLY.
Refer to your test
strip vial label for
corresponding range.
[4] Watered-down control solution.
[5] Meter malfunction.
[6] Control solution test done outside the system operating
temperature of 10°C to 40°C.
[7] Failure to shake the control solution bottle vigorously
before using.
[8] Failure to discard the fi rst drop of control solution and to
wipe the bottle tip clean.
CAUTION: Results from control solution tests do not refl ect
your blood glucose level. The control solution range is a
target range for control solution only. It is not a target
range for your blood glucose level.
21