Datron 4708 User Handbook Manual page 97

Autocal multifunction standard
Table of Contents

Advertisement

Stage
4:
To
Verify
the
1mV
Range
Full
Range
Output
a
.
Ensure
that
the
DMM
is
still
connected
to
the
4708
terminals
as
shown
in
Fig.
2.
b.
Ensure
that
the
4708
is
set to
the
10m
V
Range.
Calculate the value
'V(lOm)
x C.
Set
the
OUTPUT
Display
to this
value.
c
.
Set
the
DMM
to
measure
Im
V.
d
.
Allow
the
output
to
settle.
Note
the
DMM
reading
as
'V(lt)'.
e
.
Set
the
4708
to
its
Im
V
Range.
Adjust
4708
OUTPUTti
keys
to
give a
settled
DMM
reading of
V(lt).
Note
the
4708
OUTPUT
Display
setting
as 'V(lm)'.
Record
this
setting
in
the
last
column
of
RSI
table 4(b).
Stage 5
Verification
Against
Limits
For
Validity
Tolerance Limit
Calculations
Refer
to
Appendix
1
page 7-Al-l
para
1.
a.
Sum
the Verification setup uncertainties
and
record
in
the
Us
column
of
RSI
table
2(b) as voltage deviations. (Refer
to
Appendix
2).
b
.
Calculate the
Lower
and Upper
Validity
Tolerance
Limits:
Use
Calculation
A
if
the
4708 was
last
calibrated
by
Datron.
Use
Calculation
B
if
the
4708 was
last
calibrated against
the
standards
being used
for
verification.
c
.
Each 4708
mV
Full
Range
is
verified
if
the
4708
OUTPUT
Display
settings
recorded
in
RSI
Table
2(b) are
at
or
between
the
corresponding
Validity
Tolerance
Limits.
Calculate
in
5.5
digits
resolution
V(lOm)
X
C
=
V(lt)
is
a
transfer
value
V(lt)
=
Record
in
4.5
digits
resolution
V(lm) =
7-22