Datron 4708 User Handbook Manual page 101

Autocal multifunction standard
Table of Contents

Advertisement

Resistance
Verification
(lOQ-lOOMD
Fixed Values)
Summary
of
Verification
Procedures
a,
4-Wire Values
The 4708
is
set to
'Remote
Sense'.
Its
four output
terminals
are
connected
to
a
suitable
DMM,
configured
in
'4-wire Resistance'
mode.
Each
standard
resistor
is
thus
measured
at
its
internal
4-
wire
terminals,
and
the
DMM
reading
is
recorded.
The
calibrated
value present
on
the
OUTPUT
Display
is
also
recorded; the
specification tolerances
and accumulated
calibration uncertainties are
added.
The
measured
value
is
checked
against
this result.
The
procedure
is
detailed
on page
7-28.
b.
2-Wire Values
Remote
Sense
is
deselected.
The 4708
Hi and
Lo
terminals are
connected
to
a
suitable
DMM,
again
configured
in
4-wire
resistance
mode. Each
standard
resistor
is
thus
measured
at
the
2-wire
terminals
of
the
4708, and
the
reading
is
recorded.
The
difference
between
the
2-wire
and
4-wire
measured
values
is
also recorded;
and checked
against the specification
tolerances.
The
2-wire
zero values (including
internal
wiring) are
similarly
checked
against
specification tolerances.
The
procedure
is
detailed
on page
7-29.
Interconnections
Interconnection
instructions in
this
section are necessarily
simple
and
basic,
and
are
mainly
intended
to
show
connec-
tions
to
the
4708.
It is
recognized
that
they
may
need
to
be
adapted
to
meet
an
individual
user's
requirements.
It
is
assumed
that
users
will
possess
knowledge
of
the
operation
and
use of
standards
equipment
such
as
ihatmentioned above.
Equipment Requirements
A
DMM
of
sufficient
accuracy,
used
to
measure
resistance.
Example:
A
Datron 1281
or 1081.
Alternatively:
A
ratiometric
resistance-measuring device of
suitable
accuracy.
Preparation
Before
attempting
any
verification
ensure
that the
follow-
ing steps
have been
carried
out.
1
.
Turn on
the
instrument
to
be
checked and
allow
at least
2
hours
warm-up
in
the specified
environment.
2.
Cancel
any
MODE
keys,
ensure
OUTPUT
is
OFF
and
check
that
'cal'
is
not present
on
the
MODE
display.
(If
'cal' is
present, turn the Calibration
Keyswitch on
the
Rear
Panel
to
its
RUN
position).
3
.
Before connecting and
operating
any equipment,
consult
the
manufacturers'
handbooks.
4.
Self-Test:
Press
the Test'
key
to
carry
out
the
test
routine
described
in
Section
4.
Terminate
the
test
routine.
5
.
If
'Spec'
Mode
is
required,
select either
24hr,
90dy
or
1
yr
on
the
Rear
Panel
CALIBRATION INTERVAL
Switch.
When
received
from
Datron,
90dy
is
the
relevant
selection.
The
report sheets
hold
figures
for
this interval.
7-26