Datron 4708 User Handbook Manual page 76

Autocal multifunction standard
Table of Contents

Advertisement

RESISTANCE
(OPTION
30)
Accuracy
(2-wire
or 4-wire
sense)
Range
(1)
Stability
24 hour
±(ppm
OUTPUT+
ppm
FS)
(1)(2)
*
Accuracy
Relative
to
Calibration
Standards
±(ppm
OUTPUT
+
ppm
FS)(2)
Calibration
Uncertainty
(ppm)
Temperature
Coeff.
(±ppm
OUTPUT/°C)
13"C- 18°C
28°C
-
33°C
Specified
Current
(Is)
Maximum
Current
(Im)
Additional
Uncertainty
for
Is<I<Im
(ppm)
24
hour
23
''C
+
rC
90 days
23°C±1°C
1
year
23°C
±
5°C
lo.oooooon
2
4
10
25
10
6
10mA
100mA
(lOx
102).P
loo.ooooon
1
1.5
3
9
5
2
10mA
25mA
(8.5 x
103).F
1
.OOOOOOOk
1
1.5
3
9
5
2
1mA
10mA
(8.5 X
10^).F
lO.OOOOOOk
1
1.5
3
9
4
2
lOOpA
2.5mA
(8.5 X
10*).F
100.00000k
1
1.5
3
10
6
2
lOOpA
1mA
(8.5
X 10®).F
1.0000000M
2
4
10
25
12
6
lOpA
lOOpA
(lOx
10^).F
10.000000M
2
10
25
50
17
10
IpA
lOpA
(15x
10»).P
100.00000M
3
15
30
70
50
20
IpA
lOpA
(15x
10«).F
*
For 2-Wire Sensing:
24
hours
stability
24
hours accuracy
90 days
accuracy
}
add
±0.1
Q
1
years accuracy
add ±0.2Q
Notes:
(1)
Range
figures
are nominal,
actual calibrated
values are
displayed.
(2)
FS
=
2
X
Range.
Other
Specifications
Connection
Selectable 2 or 4-wire connection
to resistors.
2-wire
displayed value includes
internal
lead resistance.
Guarding
Selectable
remote
or
local
guard
connection.
Protection
All
resistors
fuse-protected
to
max
applied voltage
of
120V
RMS.
6-12

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Datron 4708

Table of Contents