Mask And Air Tubing Settings - ResMed VPAP ST-A User Manual

Noninvasive ventilator and heated humidifier
Hide thumbs Also See for VPAP ST-A:
Table of Contents

Advertisement

Mask and air tubing settings

Use the following settings below for each mask type:
Mask type
Full Face
Pillows
Nasal
Pediatric
Notes:
For more information on assembling the mask see the mask user guide.
For a complete list of recommended masks and their settings go to www.resmed.com on the
Products page under Service & Support. If you do not have internet access, please contact your
ResMed representative.
The VPAP ST-A is compatible with the following air tubing:
Air tubing
ClimateLine
ClimateLine
MAX
SlimLine:
Standard
3 m
Notes:
When using the SlimLine or ClimateLine above 20 cm H
not be reached if used with an antibacterial filter. The device performance must be checked prior to
prescribing the SlimLine for use with an antibacterial filter.
The ClimateLine/ClimateLine
Settings
Full Face
Pillows
Nasal (for Ultra Mirage mask, use 'Nasal Ultra')
Pediatric
Specifications
Heated
Length: 2 m
Inner diameter: 15 mm
Heated
Length: 1.9 m
Inner diameter: 19 mm
Length: 1.8 m
Inner diameter: 15 mm
Length: 2 m
Inner diameter: 19 mm
Length: 3 m
Inner diameter: 19 mm
are designed only for use with the H5i.
MAX
Settings
Automatically detected
Automatically detected
If using the SlimLine, Standard or 3 m
air tubing, adjust the tube setting in the
Patient Setup or Clinical Setup menus.
O, the device optimum performance may
2
15
English

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

H5i