Setting The Power Line Polarity; Setting The Single Fault Mode - Kikusui TOS3200 Interface Manual

Leakage current tester
Hide thumbs Also See for TOS3200:
Table of Contents

Advertisement

Setting the Power Line Polarity

TC:POL / PCC:POL
Sets the polarity (POL) of the power line supplied to the EUT. If you set the connection desti-
nation of the measurement terminal to between the enclosure and power line, this command
will be invalid.
Command [SENSe:]TC:POLarity {NORMal|REVersed}
[SENSe:]TC:POLarity?
[SENSe:]PCC:POLarity {NORMal|REVersed}
[SENSe:]PCC:POLarity?
Parameter Value:
Response
Returns the polarity of the power line in the <character> form. If the connection destination
of the measurement terminal is set to between the enclosure and power line, NA is returned.

Setting the Single Fault Mode

TC:COND / PCC:COND
Sets the single fault mode (COND). If you set the connection destination of the measurement
terminal to between the enclosure and power line, this command will be invalid.
Command [SENSe:]TC:CONDition {NORMal|FLTNEU|FLTPE}
[SENSe:]TC:CONDition?
[SENSe:]PCC:CONDition {NORMal|FLTNEU}
[SENSe:]PCC:CONDition?
Parameter Value:
Response
Returns the single fault mode in the <character> form. If the connection destination of the
measurement terminal is set to between the enclosure and power line, NA is returned.
TOS3200
Setting the TC Measurement and PCC Measurement
NORM
Normal phase connection (default)
REV
Reverse phase connection
NORMal
Normal condition (default)
FLTNEU
Power line (neutral) disconnected condition
FLTPE
Earth line disconnected condition (TC Measurement Mode only)
* RST
RST
* * RCL
RCL
3
* RST
RST
* * RCL
RCL
25