Power On and Off

To turn on the Product, push and hold  for 2 seconds. A start-up screen
shows on the display and an indicator bar shows the status, see Figure 1. The
indicator bar increases on power up and decreases on power down. After the
start-up screen, the Product is ready to use. To turn off the Product, push and
hold  for 2 seconds.
The LCD backlight (VT04A and VT04) turns off to save battery power if a button
is not pushed for more than 2 minutes. You can push any button to turn on the
backlight before the auto off time is exceeded. This feature is disabled in Alarm
mode.
The Auto Off (VT04A and VT04) feature turns off the Product after a selected
time interval. The default setting is 10 minutes and is user-selectable as 5, 10,
15, or 20 minutes of inactivity. This feature is disabled in Alarm mode.
Figure 1. Start-Up Screen and Status Indicator
For the first time use, or when the batteries are removed for more than a few
hours, the Date and Time menu opens. See page 19 for more information
about how to set the date and time.
All visual IR thermometers need sufficient warm-up time for the most
accurate temperature measurements. This time can often vary by
model and by environmental conditions. Although most visual IR
thermometers are fully warmed up in 3 to 5 minutes, it is always best
to wait a minimum of 10 minutes if the most accurate temperature
measurement is very important to your application. When you move a
visual IR thermometer between environments with large differences in
ambient temperature, more adjustment time can be required.
VT02
Note
Visual IR Thermometer
Power On and Off
VT04A, VT04
hak03.eps
5