Dräger Alcotest 6820 med Instructions For Use Manual page 35

Table of Contents

Advertisement

Available languages

Available languages

Waiting time until device is ready to use
1. Measurement
after a previous measurement of:(breath alcohol concentration
specified)
0.25 mg/L
0.50 mg/L
1.00 mg/L
> 1.5 mg/L
after a previous measurement of:(blood alcohol concentration specified)
0.50 ‰
1.00 ‰
2.00 ‰
> 3.00 ‰
Values for other units of measurement can be calculated accordingly
(see section 11 on page 34).
Calibration
every 6 months
Safety inspection
every 6 months
Functional test/ Quality assurance
Recommended daily. Quality assurance must be performed in
accordance with §4a MPBetreibV (German act on the installation,
operation and use of medical devices) and "RiliBÄK zur
Qualitätssicherung laboratoriumsmedizinischer Untersuchungen"
(Guideline of the German Medical Association on quality assurance for
medical laboratory examinations), part B1, 2.1.1.
Dräger Alcotest 6820 med
approx. 2 s
20 s
40 s
80 s
120 s
20 s
40 s
80 s
120 s
Size
Dimensions
Weight
Power supply
Batteries
Rechargeable batteries
Interface
USB
Classification per EC Directive 93/42/EEC
Measuring instrument
Mouthpieces
Protection class
IP 54
UMDNS-Code
17-475
DIMDI registration number
DE/CA01/102/040
CE mark
Electromagnetic Compatibility Directive (2014/30/EC)
The device meets the requirements of EN15964. It can be configured
according to the standard.
Technical data
65 mm x 147 mm x 39 mm (W x H x D)
approx. 260 g
2 pcs. 1.5 V Alkaline (LR6, AA)
approx. 1.500 measurements
2 pcs. 1.2 V NiMH (LR6, AA)
for data transfer to PC
Class I with a measuring function
Class I