Power Fail Condition If a Power Fail condition arises and no battery is
To Disconnect the
Tubeset
Transport Mode To transport a patient using the Auto Logic mattress,
Shut Down Power the pump off by pressing the Run/Standby button
The maximum duration of Autofirm mode is 30
minutes.
Following the end of Autofirm the previous pump mode
is restored.
connected, disconnect the tubeset from the pump. This
will put the mattress into Transport mode and will
support the patient for up to 12 hours. Once power is
resumed, re-connect the tubeset to the pump.
To disconnect the tubeset at any time, push the button
down and pull the tubeset connector away from the
pump.
This puts the mattress into Transport mode but does not
deflate the mattress. To deflate the mattress
Refer to page 18.
disconnect the tubeset from the pump. This
automatically switches the mattress into Transport
mode.
The patient will remain supported by the mattress for up
to 12 hours.
To resume normal operation, simply reconnect the
tubeset and run the pump.
There is no Transport mode on the Aura Logic seat
cushion.
for at least two seconds to select Standby (LED amber).
If you need to completely isolate the pump from the
mains, remove the plug from the mains power socket.
If running from the battery pack, remove it from the
pump.