User Setup

7.4 mA outputs

7.4.1 Outputs 0 to 5

Attribute
88
Choose the attribute that triggers the analog output:
Measurement channel
Temperature
Current (mV)
Note: The attribute variable cannot be applied to Output 0, which is reserved for the
continuous live output signal.
Type Choose either a 0-20 mA or 4-20 mA analog output
Select between linear, dual or logarithm (see graph in
Mode
parameter is only selectable if the attribute is set to a measurement channel.
Low The value corresponding to the low end of the scale
Middle The value corresponding to the mid-point of the scale (only available in dual mode)
High The value corresponding to the high end of the scale
Select the mA OUTPUTS option to set up the parameters for
all the analog outputs.
From the list available, select the mA output you wish to set.
Table 8 Analog output parameters
Description
Figure 23 on page
89). This