Specifications - DeVilbiss 525DS Service Manual

5-liter compact oxygen concentrator
Hide thumbs Also See for 525DS:
Table of Contents

Advertisement

catalog number
Delivery rate (Lower delivery rates
available for low flow applications)
Maximum recommended Flow
outlet Pressure
electrical rating
operating voltage range
oxygen Percentage
operating altitude
(tested at 21°C only)
0-1500 M (0-4921 ft)
1500-4000 M (4921-13123 ft)
operating environment range*
5°C to 40°C, humidity range
of 10% to 95%
Power consumption
Weight
Sound level (ISO 8359:1996 from front)
Dimensions
Pressure relief valve
operating System
The visible "low oxygen" indicator
will activate at the following level
Storage conditions
equipment class and Type
approval Body
and
Safety Standard
eMc compliance To:
Typical accumulator Pressure at
Maximum recommended Flow
@ Sea Level
@ 750 M (2460 Ft)
@ 1500 M (4921 Ft)
*noTe
–TheOSDperformanceat5°Cto40°C,95%R.H.throughvoltagerangeonthe525DSverifiedat670m.
noTe
–Turn-downmodemayceaseat0.5LPMifalowoutputflowmeterisnotbeingused.
Specifications subject to change without notice
DeclaraTion oF conForMiTy
Manufacturer:
Address:
ProductDesignation:
CatalogNumber:
We herewith declare that the above-mentioned product complies with the requirements
of EC directive 93/42/EEC and the following:
Class:
QualitySystemStandardsApplied: ISO13485:2003
Notifiedbody:
MDD:
S P e c i F i c a T i o n S
DevilBiSS 5-liTer coMPacT concenTraTor
525DS
115V~, 60 Hz, 3.3 Amp
97-127V~, 60 Hz
310 Watts Average
275 Watts @ 2.5 LPM & below
48 dbA overall average (50Hz)
-40°C to 70°C, humidity range of 10% to 100%, including condensation
CSA
CAN/CSA-C22.2 No. 601.1-M90
Sunrise Medical dba DeVilbiss Healthcare
Respiratory Products Division
100 DeVilbiss Drive
Somerset, PA 15501-2125 USA
Oxygen Concentrator
525KS
IIa, Rule 2
TÜV NORD
Annex II Applied
1 to 5 LPM
5 LPM
8.5 psig (58.6 kPa)
0,5-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at nominal voltage only:
No degradation in performance expected.
No degradation in performance
across the operating
voltage range.
230V / 50 Hz - 312 Watts Average
230V / 50 Hz - 296 Watts Average @ 2.5 LPM & below
230V / 60 Hz - 387 Watts Average
230V / 60 Hz - 369 Watts Average @ 2.5 LPM & below
36 lbs. (16.3 Kilograms)
24.5"H x 13.5"W x 12"D (62.2 x 34.2 x 30.4 cm)
40 psig±5psig (276 kPa±34.5 kPa)
Time Cycle / Pressure Swing
84% ± 2%
At less than 60%, the red "service required" light will activate.)
Class II Equipment Double Insulated;
EN60601-1-2
22-27 psig (150-190 kPa)
20-25 psig (138-172 kPa)
18-22 psig (120-145 kPa)
SafetyStandardsApplied:
EMCComplianceTo:
AuthorizedRepresentative:
0044
43
525kS
220-230 V~, 50 Hz, 1.55 Amp
230 V~, 60 Hz, 1.9 Amp
187-255 V~, 50 Hz
195-255 V~, 60 Hz
40
dbA overall average (50Hz)
Type B Applied Part
TUV
EN 60601+A1+A2
ISO8359: 1996
EN 60601+A1+A2
ISO8359: 1996
EN 60601-1-2
Sunrise Medical Ltd.
Sunrise Business Park
High Street
Wollaston, West Midlands DY8 4PS
ENGLAND
44-138-444-6688